Talk:Hydrogenion flux: Difference between revisions

From Bioblast
mNo edit summary
No edit summary
 
(50 intermediate revisions by 3 users not shown)
Line 1: Line 1:
{{MitoPedia
__TOC__
|abbr=''J''<sub>H</sub><small>+</small>
|description=Volume-specific '''proton flux''' is measured in a closed system as the time derivative of proton concentration, expressed in units [pmol·s<sup>-1</sup>·mL<sup>-1</sup>]. Proton flux can be measured in an open system at steady state, when any acidification of the medium is compensated by external supply of an equivalent amount of base. The extracellular acidification rate (ECAR) is the change of pH in the incubation medium over time, which is zero at steady state. Volume-specific proton flux is comparable to volume-specific [[oxygen flux]] [pmol·s<sup>-1</sup>·mL<sup>-1</sup>], which is the (negative) time derivative of oxygen concentration measured in a closed system, corrected for instrumental and chemical background.


[[pH]] is the negative logarithm of proton activity. Therefore, ECAR is of interest in relation to acidification issues in the incubation buffer or culture medium. The physiologically relevant metabolic proton flux, however, must not be confused with ECAR.
== For critical evaluation ==
|info=[[Gnaiger 2014 MitoPathways]]
:::: The measurement of H<sup>+</sup> flux alone is not sufficient to determine, if the origin of H<sup>+</sup> is the glycolysis or other sources. For example, the carbon dioxide formed during the mitochondrial respiration acts as a net source of H<sup>+</sup> into the media and as consequence has to be taken into account. During the oxidation of the glucose, we have two main metabolic pathways involved and both have a net effect over the H<sup>+</sup> flux:
}}
:::: H2CO3 is not fully dissociated. Therefore, there is not a simple and constant stoichiometry between bicarbonate and H<sup>+</sup> production.
__TOC__
== Proton flux versus glycolytic flux ==
::::# Measured changes in pH over time (ECAR) must be transformed from the logarithmic scale to the linear scale of proton flux.
::::# Measurement of extracellular proton production and glycolytic flux are related under specifically controlled conditions. Such conditions must be carefully evaluated, may require modifications of protocols, and must be corrected for acid-base reactions unrelated to glycolytic flux.
::::# The measurement of proton flux alone is not sufficient to determine if the origin of the protons is the glycolysis or other sources. For example, the carbon dioxide formed during the mitochondrial respiration acts as a net donor of protons into the media and as consequence has to be taken into account. During the oxidation of the glucose, we have two main metabolic pathways involved and both have a net effect over the proton flux:


[[File:Proton production Metabolic pathways.png|center|400ppx]]
[[File:Proton production Metabolic pathways.png|center|400ppx]]
:[[File:Lactate.png|center|400ppx]]
:[[File:Lactate.png|center|400ppx]]
:[[File:Carbon dioxide.png|center|400ppx]]
:[[File:Carbon dioxide.png|center|400ppx]]
:::: As we can observe, the production of protons by the oxidative phosphorylation is three times higher than the one produced by the glycolysis per molecule of glucose. However, the chemical rate of production could be determinant to determine which is the main source of protons in our sample under specific conditions.
* We have also to take into account the pka for the point of equilibrium of the most common weak acids that will be formed during both processes:


== Proton flux versus ECAR ==
:::: The extracellular acidification rate (ECAR) is the change of pH in the incubation medium over time and can only be measured in a closed system (why?). pH is the negative decadic logarithm of proton activity which is, in diluted solutions, in close approximation to the negative decadic logarithm of proton concentration.
:::: Thus, measured changes in pH over time (ECAR) must be transformed from the logarithmic to the linear scale to obtain proton flux.
Therefore, ECAR is of interest in relation to acidification issues in the incubation buffer or culture medium but must not be confused with the physiologically relevant metabolic proton flux.


== H<sup>+</sup> flux and glycolysis ==
:::: Measurement of proton flux and glycolysis are related under specifically controlled conditions. Such conditions must be carefully evaluated, may require modifications of protocols, must be corrected for acid-base reactions unrelated to glycolysis and thus need data analysis beyond reporting changes of pH.
:::: Glycolysis is the degradation of glucose to pyruvate. Depending on the subsequent metabolism of pyruvate, glycolysis is indirectly related to H<sup>+</sup> flux. Pyruvate can either be converted to lactate, catalyzed by [[lactate dehydrogenase]] in the cytosol, or converted to Acetyl-CoA catalysed by [[pyruvate dehydrogenase]]), feeding into the [[Tricarboxylic acid cycle | TCA cycle]] (mitochondria). The catabolism of pyruvate can have an impact on proton flux as illustrated by the following equations:


== Additional resources ==
» O2k-Manual: [[MiPNet23.15 O2k-pH ISE-Module]]


» O2k-SOP: [[MiPNet08.16 pH calibration]]
»{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware
}}


{{Technical support integrated}}
[[File:Proton_production.png]]
== O2k signal and output ==
:::# [[O2k signals and output#Signal of the O2k and add-on modules |O2k signal]]: The [[O2k-pH ISE-Module]] is operated through the pX channel of the O2k, with electric potential (volt [V]) as the primary and raw signal
:::# [[O2k signals and output#O2k output |O2k output]]: type I and II




== pH changes versus glycolytic flux ==
:::: Measurement of extracellular proton production and glycolytic flux are related under specifically conrolled conditions. Such conditions must be carefully evaluated, may require modifications of protocols, and need data analysis beyond reporting changes of pH.


::::* The extracellular acidification rate (ECAR) is the change of pH over time, which may be of interest in relation to acidification problems in a culture medium or incubation buffer. [[pH]] is the negative logarithm of proton activity. Comparable to volume-specific [[oxygen flux] [pmol·s<sup>-1</sup>·mL<sup>-1</sup>]], which is the (negative) time derivative of oxygen concentration measured in a closed system, volume-specific [[proton flux]] is the time derivative of proton concentration, expressed in units [pmol·s<sup>-1</sup>·mL<sup>-1</sup>]]. The physiologically relevant metabolic proton flux, therefore, must not be confused with ECAR.
::::» [[Proton flux]]


::::* Very small buffering capacity is required: To accurately measure biologically induced changes in pH, the buffering capacity of the medium has to be small. This may be addressed either by using or preparing media with a buffering capacity that is low but still sufficient to keep the pH in the desired range for a limited period of time. An alternative approach is to use buffers with very low buffering capacity and keep the pH value inside the desired limits by a [[O2k-pH_ISE-Module#pH-Stat|pH-Stat]].


As we can observe, the production of protons per molecule of glucose is three times higher in the mitochondria via dissolution of carbone dioxide  than by conversion of pyruvate to lactate in the cytosol. Furthermore, the equations illustrate that the measurement of proton flux alone is not sufficient to determine the origin of protons. Experimental settings can help to estimate the main source of proton production, e.g. by inhibition of [[Oxidative phosphorylation | OXPHOS]] (link to SUIT protocols).


== Compare measurement of pH with the pH electrode and ratiometric fluorometric methods (NextGen-O2k) ==
To accurately measure proton flux induced by biological sample, the buffering capacity of the medium has to be small but still sufficient to keep the pH in the desired range for a limited period of time. Furthermore, the buffering capacity has to be determined ([[MiPNet23.15 O2k-pH ISE-Module]]) and taken into account when proton flux is calculated from the measured changes in pH. An alternative approach is to use buffers with very low buffering capacity and keep the pH value inside the desired limits by a [[O2k-pH_ISE-Module#pH-Stat|pH-Stat]]. Here, proton flux can be calculated either by changes in pH over time (previous calculation of buffering capacity of the medium required) or by the amount of injected base via pH Stat. [[MiPNet23.15 O2k-pH ISE-Module]]
::::» [[Carboxy SNARF 1]]
::::» [[HPTS]]


== Measurement of proton flux with the O2k-pH ISE-Module ==


== O2k-Manual ==
:::: The [[Oroboros O2k]] supports the modular O2k-MultiSensor extension for recording potentiometric (voltage) signals simultaneously with the oxygen signals in both O2k-chambers. Potentiometric measurements result in a voltage signal ('''pX''') which is  typically a linear function of the logarithm of the activity  (concentration) of the substance of interest (the ''analyte''). A calibrated pH electrode displays the negative decadic logarithm of the H<sup>+</sup> ion activity (potentia hydrogenii) and thus got its name “pH electrode”.
=== Introduction and scope ===
:::: The [[Oroboros O2k]] supports the modular O2k-MultiSensor extension for recording potentiometric (voltage) signals simultaneously with the oxygen signals in both O2k-chambers. A new pH probe system is described here,  consisting of separate reference and measuring electrodes.  


::: '''Definition of terms'''
== Applications  ==


::::* '''pH/ISE''':  The potentiometric channels may be used with different kinds of pH / ISE (ion selective electrode) modules, (i) with the pH / reference electrode  module described here, or (ii) with an ISCE – an ion selective combination electrode, combining reference and measuring electrode into one sensor (pH combination electrodes, other ion selective electrodes; see [[O2k-TPP%2B_ISE-Module]]). The O2k not only includes the two  potentiometric channels, but supports two additional amperometric (current) channels for the [[O2k-Fluo_LED2-Module]] or amperometric sensors (NO, H<sub>2</sub>O<sub>2</sub>, etc).
:::: The majority of novel applications will address aerobic or anaerobic glycolysis in intact cells, using the measurement of proton flux as an indirect but continuous record of lactate production and corresponding acidification of the medium, while simultaneously monitoring oxygen concentration and oxygen consumption. For this application, specific experimental settings are required to leave lactic acid production as the dominant mechanism of acidification.


::::* '''pX''':  Potentiometric measurements result in a voltage signal which is  typically a linear function of the logarithm of the activity  (concentration) of the substance of interest (the ''analyte''). A  calibrated pH electrode displays the negative decadic logarithm of the H<sup>+</sup> ion activity (potentia hydrogenii) and thus got its name “pH electrode”. By analogy, an ISE may be used to measure pTPP, pCa, etc., hence the general term “pX” is used to denote the signal from such an ISE.
:::: The pH electrode in the O2k can also be used in conjunction with a study of mitochondrial permeability transition (e.g. [[SE_Lund_Elmer E]]).


:::: For simultaneous measurement of O<sub>2</sub> and pH, we refer to the classical literature on bioenergetics and the discovery of the chemiosmotic coupling mechanism, the quantification of H<sup>+</sup>/O<sub>2</sub> stoichiometric ratios for proton pumping (Peter Mitchell).




:::: '''Conditioning:''' Optimum response time will be obtained after the probe has been exposed to two buffer solutions. Place a pH 4 buffer or equivalent in a  beaker and a pH7 buffer or equivalent in a second beaker. Hold the pH  electrode and reference electrode together and touch the pH 4 buffer surface, allowing 15-20 seconds of equilibration. Rinse the two electrodes with distilled water and then touch the pH 7 buffer surface in the same manner. Repeat several times.


:::: '''Handling:''' Be careful not to apply pressure against, or to shock, the inner glass capillary tube.
== O2k signal and output ==
:::# [[O2k signals and output#Signal of the O2k and add-on modules |O2k signal]]: The [[O2k-pH ISE-Module]] is operated through the pX channel of the O2k, with electric potential (volt [V]) as the primary and raw signal
:::# [[O2k signals and output#O2k output |O2k output]]: type I and II


:::: '''Cleaning:''' When using the electrode in solutions containing protein, the  pH electrode and reference electrode should be soaked in enzyme cleaning  solution such as Terg-a-zyme (Alconox Inc), or a chromic/sulfuric acid  glass cleaning solution after each use for a few minutes to remove the protein from the glass and the reference junction. This will prolong the lifetime of the electrode.


:::: '''Storage: Always clean the electrode before storing.


::::* '''Long-term storage (over 2 weeks):''' Place the electrode into its original container in the same condition in which your received it. Usually this  means moistening with distilled water the sponge located in the bottom of the protective glass tube.
== Compare measurement of pH with the pH electrode and ratiometric fluorometric methods ==
::::» [[Carboxy SNARF 1]]
::::» [[HPTS]]


::::* '''Short-term storage:''' The electrode can be left in acidic pH buffer solution, e.g. pH 4.
{{Keywords: pH}}


:::: '''Troubleshooting:''' If possible try to locate the problem either at the measuring pH or at the reference electrode by switching electrodes. If you have only one reference electrode you can switch to a spare glass barrel for  diagnostic purposes. The following text assumes that the problem was located on the reference electrode.
== References ==


::::* '''Little or no response:''' Inspect the electrode for visible cracks (usually occurring at the tip of the electrode). If any exists, the electrode is defective and must be replaced. The slightest crack in or around the tip of the electrode may cause the electrode to read about the same in all solutions.
::::* [[MiPNet08.16 pH calibration]]
::::* [[MiPNet15.03 O2k-MultiSensor-ISE]]


::::* '''Response pegs off-scale:''' (i) Check the pX gain setting. (ii) Visually inspect the electrode for a broken bulb.
::::* [[MiPNet23.15 O2k-pH ISE-Module]]


::::* '''Sluggish response:''' If the electrode becomes sluggish in responding to changes  in pH, the response time can be improved using the following procedure:
::::* [[MiPNet24.06 Oxygen flux analysis - DatLab 7.4]]


::::# Clean the electrode as described above.
::::* [[MiPNet12.10 TIP2k-manual]]
::::# Soak the electrode in 0.1 N HCl for 5 minutes, followed by soaking in  0.1 N NaOH for 5 minutes. After repeating several times, rinse the  electrode thoroughly with distilled water. The electrode can then be  calibrated in the usual manner.


::::* [[MiPNet15.08 TPP electrode]]


=== Reference electrode: assembly, storage and maintenance ===
::::* [[MiPNet22.11 O2k-FluoRespirometer manual]]   


:::*-See: [[MiPNet15.03 O2k-MultiSensor-ISE]] in section 2.5.Reference electrode: assembly, storage and maintenance
=== O2k-MultiSensor System ===
:::: Applications of the O2k with one or two potentiometric electrodes (ISE or pH) require an O2k Series D and higher, or a previous O2k-Series with electronic MultiSensor upgrading. In addition to the two polarographic (amperometric) oxygen channels, the O2k provides two electronic channels for potentiometric (voltage)  measurements with ISE or pH (O2k Series B upwards). The Multisensor  function of O2k Series D (and higher) is extended further with two  additional amperometric channels (current measurement; for NO, etc.). Please see [[O2k-Main Unit#O2k-Series]] for how to determine the series  of an oxygraph.


==== O2k-FluoRespirometer ====
:::: '''Connect:''' In  O2k Series D and higher, pH and reference electrodes are directly connected to the plugs on the front side of the O2k. Insert the  connector of the pH electrode into the BNC plug labelled “pX” and the  connector of the reference electrode to the 2 mm pin plug labelled  “Ref”. See [[MiPNet22.11 O2k-FluoRespirometer manual]].


:::: '''Gain:''' The gain of the pX channel can be selected in the DatLab software in  the [[O2k configuration]] window. For measurements with the OROBOROS  pH system, a gain of 20 is suggested. Usually it will not be necessary  to change the gain for pH work.
»{{MitoPedia O2k and high-resolution respirometry


==== O2k Series B and C, pX upgrade ====
|mitopedia O2k and high-resolution respirometry=O2k hardware
::::» [[Talk:MiPNet23.12 O2k-pH ISE-Module]]
 
 
=== Operation instructions  ===
==== Insertion of electrodes: bubble-free filling of the chambers ====
:::*-See: [[MiPNet15.03 O2k-MultiSensor-ISE]] in section 4.1 Insert ISE
 
==== Volume calibration with MultiSensor stoppers ====
:::* ''-See:'' [[MiPNet19.18A O2k-Series G: Start]].
 
::::# Add to the dry  O2k-chamber containing the stirrer bar a water volume accounting for the  final chamber volume (2 mL) plus the additional dead volume in the  capillary and spaces between electrodes and inlets. For the OROBOROS pH  Assembly (ion selective electrode + reference electrode), this  additional volume is approximately 0.14 mL. Therefore, the necessary  volume to calibrate a chamber volume of 2 mL with the Oroboros pH system  is 2.14 mL.
::::# Start stirring, cover the chamber with a Perspex  cover or a loosely placed stopper, and wait for equilibration. To avoid  creating bubbles during the calibration process it is very important to  allow for full thermal equilibration of the liquid in the chamber.  Continue with volume calibration only after reaching the conditions for  oxygen calibration at air saturation (stable temperature and Peltier  power, near-zero uncorrected oxygen flux (±1 pmol∙s<sup>­1</sup>∙mL<sup>­1</sup>).
 
::::# Afterwards follow instructions written in [[MiPNet15.03 O2k-MultiSensor-ISE]] in section 4.1 Insert ISE
 
==== Handling during an experiment ====
:::*-See: [[MiPNet15.03 O2k-MultiSensor-ISE]] in section 4.3. Experiment
 
==== Instrumental background oxygen flux ====
:::*-See: [[MiPNet15.03 O2k-MultiSensor-ISE]] in section 4.3. Instrumental background oxygen flux
 
==== pH calibration ====
::::» [http://wiki.oroboros.at/images/a/ab/MiPNet19.18D_O2k-Calibration.pdf Calibration] and [[MiPNet08.16 pH calibration]].
::::» [[PH calibration buffers]]
 
==== Performance, trouble shooting, and electrode lifetime ====
:::: '''Performance criteria '''can be assessed by two tests:
 
:::: '''1. Calibrations''' must be reproducible: After performing a two point calibration,  reinserting the electrodes in the first used calibration buffer should  give the correct calibrated pH value. The slope and intercept of the calibration can be copied to a spreadsheet file ([[File:PH-Calibration-List.xls]]) to obtain a track record..
 
:::: '''2.Drift:''' The drift in the medium without biological sample hast to be small  as compared to pH changes expected from the sample. The drift will be  different in different media (buffering capacity), so for each used  system the necessary experience has to be gained. The drift is the first  time derivative of the pH signal, so in DatLab it will be shown as “pX  Slope”. After proper calibration the unit of this plot will be mpH/s.
 
:::: '''Trouble shooting: '''If the required performance criteria are not reached, the following steps should be tested:
 
::::# Set the polarisation voltage of the O2 sensor to 0 V in the O2k Control  window. Observe any effects on the pX raw signal. A tiny potential jump  is acceptable. If a drift in the pX signal is either increased or  reduced by this test or an extreme jump in the signal observed, the  membrane of the polarographic oxygen sensor (POS) should be replaced.  Reset the polarisation voltage to 800 mV after the test.
::::# If you  have two pH electrodes: Locate the problem to either the  reference  electrode or the pH electrode by switching either only the pH  or only  the reference electrode between chambers.
::::# Follow the specific trouble shooting procedures for the reference electrode or the pH electrode described above.
::::# '''pH electrode lifetime: '''All glass pH electrodes do have a limited  lifetime due to aging of the glass membrane. After 1.5 to 2 years a loss  of performance will occur even without use. Therefore, we strongly recommend to obtain pH electrodes only when required for a defined project and not buy the electrodes ahead of need.
 
=== Measuring proton production ===
==== From the change of the pH signal ====
 
:::: To calculate the actual proton flow in the system from the observed  change  in the pH signal the buffering capacity of the medium has to be  determination of before the introduction of sample. This is best achieved by setting up the [[TIP]] with HCl in the syringes and simulate the expected proton production of the sample by setting an appropriate flow rate for the TIP. In the simplest implementation just one flow rate is used for about 5 minutes. From the first time derivative of the pH signal the buffering capacity can be calculated. Afterwards the TIP has to be re-fitted with syringes containing KOH (NaOH) for using the pH-Stat during the actual biological experiment. The observed rates of pH changes during the biological experiment (pX slope/ first time derivative of pH signal) can then be directly converted to Proton flow values using the buffering capacity determined before. This is shown in the spreadsheet file "pH Stat Template Buffer Capacity_one_point" In a more sophisticated approach a multiple point calibration is done: Several proton flows are simulated before the experiment, a linear regression between set proton flow and observed pH slope is done and the regression parameters are used to calculate proton flows from the pX slopes observed during the biological experiment.
 
:::: These approaches assume a linear relationship between pH change and introduced protons. This is an approximation that is only valid for very  small pH changes. In other words the buffering capacity has to be constant during the entire experiment.
 
:::: For calculating proton production rates following this approach directly in DatLab, see below.
 
==== From the amount of injected base ====
 
:::: The limitations mentioned above can be overcome by using the amount of  base  necessary in the pH-Stat approach to hold the pH value constant.  If the  pH values at the beginning and at the end of a time interval are identical then the proton floe during this time interval can be  directly  calculated from the amount of base injected to keep the pH  value constant. While this method does not assume a constant buffering  capacity during the entire experiment there are some drawbacks:
 
::::* Usually only the "pH-Stat-strict" approach will assure the identical pH values at the beginning and the end of a time period necessary for this  approach (This could possibly be overcome by more advanced data analysis).
::::* Only one value for proton flow for each period between injections is calculated in contrast to the continuous recording facilitated by the first mentioned method.
::::* The injected volumes  have to be read out form TIP events. This can be partially automated in a spreadsheet template ([[Media:PH Stat Template Injected Volume filled .xls|pH Stat Template Injected Volume]])  but is more tedious than the method using the buffering capacity
::::* Initial trials with simulated proton flows indicted this method to be slightly less precise than the method using the buffering capacity.
 
==== Conclusion and templates ====
 
:::: A potential compromise would be to use the method based on buffering capacity for routine calculations but check selected (late) phases of the experiment with the method based on added base volume. Thereby, significant changes in buffering capacity during the experiment should be detected. Spreadsheet (Excel) templates for both methods are available for download [[here|here]].
 
=== DatLab analysis ===
==== [[Image:Px referencelayouts.png|right|600 px]] pX reference layouts ====
:::: '''Graph  layout''': Four [[Layout for DatLab graphs|reference layouts]] are available in [[DatLab 7]] based on the recorded pX  signal:
:::::*'''01 Potentiometric'''
:::::*'''02a TPP_calibration'''
:::::*'''02b TPP_with_O2flux'''
:::::*'''02c TPP_calibrated_with_O2flux'''
:::: These layouts can be selected in [Layout / Reference layouts / O2 & pX].
 
:::: '''Reference layouts''' can be modified and saved as user-defined layouts, see [[MitoPedia: DatLab]].
 
 
:::: [[Image:O2kcontrol px.PNG|400 px|right]]
 
==== pX settings  ====
 
:::: In the '''[[O2k configuration]] window''' the pX channel can be activated and a label for the inserted pX electrode can be entered for documentation purposes.
:::: '''Gain''' and '''Offset voltage [mV]''' for the pX channel can be set in the '''[[O2k control]] window''' [F7], tab: '''Potentiometric, pX'''. The gain influences the  "pX Raw Signal” recorded in DatLab. Therefore, a gain of 1 will give the  same voltage [V] as would be measured with any multimeter between  reference and measuring electrode.
 
 
 
==== pX calibration ====
 
:::: From  the Calibration menu the '''pX Calibration window''' window [Potentiometric, pX] is opened. This feature allows for a simple two-point  linear calibration of pX (or –pX) as a function of recorded voltage,  using data ranges marked for calibration [Select marks] or known pX values, see [[MiPNet19.18D O2k-Series G and DatLab 6: Calibration| O2k-calibration]] and [[MiPNet19.18D O2k-Series G and DatLab 6: Calibration| pH calibration]].
 
 
:::: '''Calibrations  for different signal types: '''There is only one set of calibration  values for each pX channel, irrespective of the connected electrode. If a  pX channel was calibrated for a pH electrode, these values will  initially also be used to calculate the calibrated signal when the pH  electrode is exchanged for a TPP<sup>+</sup> electrode. Even  when observing only the raw (not the calibrated) signal, the time  derivative (Slope pX) will be calculated from the calibrated signal,  which might lead to confusion when the time derivative is used to access  stability or signal drift. It is, therefore, suggested to set the  calibrated signal to the raw signal whenever the raw signal is to be  used as the primary data source.
 
:::: [[Image:PX calibration window.JPG| right|500 px]] Calibration values from other files can be imported with '''Copy from file''' in the pX calibration window.
 
 
==== Calculate proton production in DatLab ====
:::: Note: Requires DatLab 5 or [[DatLab 6]]
 
:::: Proton production rates can be calculated in real-time during data acquisition. Select the menu [Flux/Slope]/[Proton Flux].
::::# Determining the buffering capacity of the medium:
::::## Calibrate the pH electrode, observe the calibrated pH signal.
::::## Fill TIP syringes with diluted acid or base.
::::## Start a slow injection  of acid or base into the media (no sample present).
::::## Place a mark on a stable region of the slope plot of the calibrated pH signal 'pX slope".
::::## Go to [Plots]/[Proton Flux].
::::## The buffering capacity is calculated by DatLab and can be used for    calculations of biological proton flow in the same file or noted down and used in subsequent experiments.
::::# Biological proton flux:
::::## Calibrate the pH electrodes.
::::## Observe the pH calibrated signal.
::::## Place marks on regions of interest on the pX slope plot. If you use  the  "pH" stat" se above to keep the pH value in desired limit make sure that you exclude times during which base was injected.
::::## Select [Flux/Slope] / [Proton Flux].
::::## Enter the buffering capacity in the appropriate field or use the    feature in the upper part of the window to calculate the buffering capacity from a calibration experiment in the same file.
::::## Press [OK].
::::## A new plot "Proton Flux" is now available in [Graph]/[Select Plots] (right at the end of the list. You can now chose to display this plot e.g. instead of the pX slope plot by selecting its check box and de-selecting the check box of the "pX slope" plot.
 
:::: Known issues: DatLab always calculates a new buffering capacity from the input in the upper part of the window and does not remember the value from previous files. Therefore, if the determination of buffering capacity was done in a different Datlab file the value has to manually entered.
 
 
== Specifications ==
 
:::: Specifications provided by OROBOROS INSTRUMENTS for quality control of pH electrodes:
::::* Drift (after 45 min stabilization, integrated over 5 minutes, 37 °C, 2 mM buffering capacity): <= 20 µpH/s.
 
== pH-Stat ==
 
:::: One approach we have developed is to use the [[TIP]] to run in a "pH stat" mode, i.e. keeping the pH constant by a feedback controlled automatic injection of base. Besides keeping the pH in the desired range this can actually be used to determine proton flow from the amount of base injected, circumventing the determination of buffering capacity, see below. The "pH-Stat" allows using very weakly buffered media (2 mM buffering substances) and might even make buffering obsolete. The feedback modus of the [[TIP]] can be used in two ways to achieve a pH-Stat modus:
 
::::* pH-Stat_strict: This program keeps the pH value strictly between user defined upper and lower limits. The difference between the upper and lower limit will determine the time between injections depending on the current proton flow. Therefore, the time between injections may vary drastically with changing proton fluxes.
 
<br>
 
::::* pH-Stat_interval: This program adjusts the pH value in certain time intervals back to the upper limit. The difference between the upper and lower limit is set extremely small but a defined minimum pause between injections of e.g. 180 s is defined. Therefore, usually a base injection will be done every 3 minutes and the pH value will oscillate between the upper limit and some (proton flux dependent) lower limit. The lower limit set in the program has no significance because the minimum pause time will not have elapsed when the lower limit is met.
 
:::: While the pH-Stat_strict is necessary to keep the pH value in a precisely defined range the pH-Stat_interval program ensures defined periods undisturbed by any injection of base. Such periods are necessary for the calculation of proton flow from the observed pH change and theoretically also for measuring respiration (however, if a 100 mM KOH the disturbance of the oxygen signal by the small amounts of KOH added was usually very small).
 
:::: TIP2k setups in the DatLab template file '''DLTemplates_pH.dlt''' and Spreadsheet (e.g. Excel) templates for determining proton form base injections are available for download: [[File:DLTemplates.dlt| here]]. Please note that of course the more straightforward calculation of proton flows from the measured pH slope is also possible while operating in pH-Stat mode!
 
== Applications  ==
 
:::: For simultaneous measurement of O2 and pH, we refer to the classical literature on bioenergetics and the discovery of the chemiosmotic coupling mechanism, the quantification of H+/O2 stoichiometric ratios for proton pumping (Peter Mitchell). Other groups (e.g. [[SE_Lund_Elmer E]]) have used the pH electrode in the O2k in conjunction with a study of mitochondrial permeability transition.
 
:::: The majority of novel applications will address the problem of aerobic glycolysis in intact cells, using the measurement of proton production as an indirect but continuous record of lactate production and corresponding acidification of the medium, while simultaneously monitoring oxygen concentration and oxygen consumption. In a well buffered culture medium, the pH change is extremely small relative to the amount of protons (lactic acid) produced, hence a low-buffering capacity medium needs to be applied. A titration of acid (lactic acid or HCl) into the low-buffering capacity medium yields the pH-dependent buffering capacity (Delta H+ added/Delta H+ measured by the pH electrode). Under various metabolic conditions, lactic acid production is the dominant mechanism causing acidification, hence the pH measurement is a good indirect indicator of aerobic glycolysis.
 
== Protocols ==
 
:::* [[SUIT-003 O2 ce D067]]
 
{{Template:SUIT text D067}}
== Demo Experiment with simulated proton flow ==
[[File:PH PS demofile 6.png|500px]]
 
 
:::: Medium: imidazole buffered medium, see above
 
::::# Calibration steps (for calculating buffering capacity): 30, 90, 150 HCl pmol/(s mL): TIP 1 mmol/L HCl, pump speed 0.06, 0.18, 0.3 µL/s
::::# Simulated proton flow 30,90,150 pmol HCl/(s mL) in pH stat mode: The pH value was held inside narrow limits by using the TIP in pH stat mode (100 mM KOH, Tip set up similar to the one included in DLTemplates_pH.dlt available from http://www.oroboros.at/index.php?id=ph-oxygen. The proton flow was simulated using a second TIP
 
 
{{Keywords: pH}}
 
== References ==
 
::::* '''O2k-Manual''': [[Media:MiPNet19.18 O2k-Core Manual.pdf|Contents: O2k-Core Manual.pdf]]
{{#ask:[[Category:Publications]] [[Instrument and method::O2k-Manual]] [[Additional label::O2k-Core]]
| mainlabel=Chapter
|?Has title=Section
|?Was published in year=Last update
|format=broadtable
|limit=500
|sort=
|order=ascending
|offset=0
}}
}}
 
»{{MitoPedia O2k and high-resolution respirometry
[[Image:Titration-Injection-microPump.jpg|150px|left|link=http://www.bioblast.at/index.php?title=TIP2k-Module]]
|mitopedia O2k and high-resolution respirometry=DatLab
::::* '''TIP2k-Manual***
::::» [[O2k-Catalogue: TIP2k]]
::::» [[O2k-Publications: TIP2k]]
{{#ask:[[Category:Publications]] [[Instrument and method::O2k-Manual]] [[Instrument and method::TIP2k]]
| mainlabel=Chapter
|?Has title=Section
|?Was published in year=Last update
|format=broadtable
|limit=500
|sort=
|order=ascending
|offset=0
}}
}}
{{Technical support integrated}}


{{Keywords: pH}}


{{MitoPedia concepts
{{MitoPedia concepts
Line 291: Line 87:
}}
}}
{{MitoPedia methods
{{MitoPedia methods
|mitopedia method=Respirometry, Fluorimetry
|mitopedia method=Respirometry
}}
}}

Latest revision as of 14:04, 5 December 2020

For critical evaluation

The measurement of H+ flux alone is not sufficient to determine, if the origin of H+ is the glycolysis or other sources. For example, the carbon dioxide formed during the mitochondrial respiration acts as a net source of H+ into the media and as consequence has to be taken into account. During the oxidation of the glucose, we have two main metabolic pathways involved and both have a net effect over the H+ flux:
H2CO3 is not fully dissociated. Therefore, there is not a simple and constant stoichiometry between bicarbonate and H+ production.
400ppx
400ppx
400ppx

Proton flux versus ECAR

The extracellular acidification rate (ECAR) is the change of pH in the incubation medium over time and can only be measured in a closed system (why?). pH is the negative decadic logarithm of proton activity which is, in diluted solutions, in close approximation to the negative decadic logarithm of proton concentration.
Thus, measured changes in pH over time (ECAR) must be transformed from the logarithmic to the linear scale to obtain proton flux.

Therefore, ECAR is of interest in relation to acidification issues in the incubation buffer or culture medium but must not be confused with the physiologically relevant metabolic proton flux.

H+ flux and glycolysis

Measurement of proton flux and glycolysis are related under specifically controlled conditions. Such conditions must be carefully evaluated, may require modifications of protocols, must be corrected for acid-base reactions unrelated to glycolysis and thus need data analysis beyond reporting changes of pH.
Glycolysis is the degradation of glucose to pyruvate. Depending on the subsequent metabolism of pyruvate, glycolysis is indirectly related to H+ flux. Pyruvate can either be converted to lactate, catalyzed by lactate dehydrogenase in the cytosol, or converted to Acetyl-CoA catalysed by pyruvate dehydrogenase), feeding into the TCA cycle (mitochondria). The catabolism of pyruvate can have an impact on proton flux as illustrated by the following equations:


Proton production.png



As we can observe, the production of protons per molecule of glucose is three times higher in the mitochondria via dissolution of carbone dioxide than by conversion of pyruvate to lactate in the cytosol. Furthermore, the equations illustrate that the measurement of proton flux alone is not sufficient to determine the origin of protons. Experimental settings can help to estimate the main source of proton production, e.g. by inhibition of OXPHOS (link to SUIT protocols).

To accurately measure proton flux induced by biological sample, the buffering capacity of the medium has to be small but still sufficient to keep the pH in the desired range for a limited period of time. Furthermore, the buffering capacity has to be determined (MiPNet23.15 O2k-pH ISE-Module) and taken into account when proton flux is calculated from the measured changes in pH. An alternative approach is to use buffers with very low buffering capacity and keep the pH value inside the desired limits by a pH-Stat. Here, proton flux can be calculated either by changes in pH over time (previous calculation of buffering capacity of the medium required) or by the amount of injected base via pH Stat. MiPNet23.15 O2k-pH ISE-Module

Measurement of proton flux with the O2k-pH ISE-Module

The Oroboros O2k supports the modular O2k-MultiSensor extension for recording potentiometric (voltage) signals simultaneously with the oxygen signals in both O2k-chambers. Potentiometric measurements result in a voltage signal (pX) which is typically a linear function of the logarithm of the activity (concentration) of the substance of interest (the analyte). A calibrated pH electrode displays the negative decadic logarithm of the H+ ion activity (potentia hydrogenii) and thus got its name “pH electrode”.

Applications

The majority of novel applications will address aerobic or anaerobic glycolysis in intact cells, using the measurement of proton flux as an indirect but continuous record of lactate production and corresponding acidification of the medium, while simultaneously monitoring oxygen concentration and oxygen consumption. For this application, specific experimental settings are required to leave lactic acid production as the dominant mechanism of acidification.
The pH electrode in the O2k can also be used in conjunction with a study of mitochondrial permeability transition (e.g. SE_Lund_Elmer E).
For simultaneous measurement of O2 and pH, we refer to the classical literature on bioenergetics and the discovery of the chemiosmotic coupling mechanism, the quantification of H+/O2 stoichiometric ratios for proton pumping (Peter Mitchell).


O2k signal and output

  1. O2k signal: The O2k-pH ISE-Module is operated through the pX channel of the O2k, with electric potential (volt [V]) as the primary and raw signal
  2. O2k output: type I and II


Compare measurement of pH with the pH electrode and ratiometric fluorometric methods

» Carboxy SNARF 1
» HPTS


Questions.jpg


Click to expand or collaps


References


» MitoPedia O2k and high-resolution respirometry: O2k hardware 

» MitoPedia O2k and high-resolution respirometry: DatLab 


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 




MitoPedia concepts: MiP concept 


MitoPedia methods: Respirometry 

Cookies help us deliver our services. By using our services, you agree to our use of cookies.