- high-resolution terminology - matching measurements at high-resolution
Oxygen flux - instrumental background
Description
Instrumental background oxygen flux (JΒ°O2) MiPNet14.06, in a respirometer is due to oxygen consumption by the POS, and oxygen diffusion into or out of the aqueous medium in the O2k-chamber. It is measured in the range of experimental oxygen levels by a standardized instrumental background test, and is a property of the instrumental system. Instrumental background correction eliminates errors by systemic flux compensation, automatically performed by DatLab.
Abbreviation: JΒ°O2
Reference: Gnaiger_2008_POS; Gnaiger_2001_RespPhysiol
Labels:
Regulation: Respiration; OXPHOS; ETS Capacity"Respiration; OXPHOS; ETS Capacity" is not in the list (Aerobic glycolysis, ADP, ATP, ATP production, AMP, Calcium, Coupling efficiency;uncoupling, Cyt c, Flux control, Inhibitor, ...) of allowed values for the "Respiration and regulation" property.
HRR: Oxygraph-2k, DatLab Software; Separate Application"DatLab Software; Separate Application" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property., Method"Method" is not in the list (Oxygraph-2k, TIP2k, O2k-Fluorometer, pH, NO, TPP, Ca, O2k-Spectrophotometer, O2k-Manual, O2k-Protocol, ...) of allowed values for the "Instrument and method" property.