Difference between revisions of "O2k-OroboPOS-Holder MitoPedia"

From Bioblast
(Redirected page to OroboPOS-Holder)
Tag: New redirect
Β 
Line 1: Line 1:
{{MitoPedia
#REDIRECT [[OroboPOS-Holder]]
|description=
The OroboPOS-Holder (POS-Holder), made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding the OroboPOS-Connector with the OroboPOS sensor head (POS) to the O2k-Chamber, and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-Chamber with the OroboPOS-Seal Tip. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.
Β 
Two units of this item are standard components mounted on the O2k-Main Unit.
|info=[[OroboPOS-Holder]]
}}
__TOC__
Β 
{{Template:Technical support integrated}}
== Assembly ==
Β 
::::* Manual: [https://wiki.oroboros.at/images/2/29/MiPNet22.11_O2k-FluoRespirometer-manual.pdf MiPNet22.11_O2k-FluoRespirometer-manual.pdf] Section 5.2.
Β 
::::* Video: [https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]
== A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber ==
Β 
:::* The OroboPOS-Holder may be damaged:
Β 
::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (''see center picture below'').Β 
::::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]].
Β 
::::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see'' pictures below).
Β 
[[File:OroboPOS-Holder002.jpg|left|200px]]
[[File:OroboPOS-Holder003.jpg|right|200px]]
[[File:OroboPOS-Holder004.jpg|centre|200px]]
Β 
* Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
* Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged.
* Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.
Β 
Β 
::::* Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: [https://www.oroboros.at/index.php/o2k-products/spares/ Spare parts & consumables]
''See'' [[Talk:O2k-OroboPOS-Holder MitoPedia|Discussion page]]
Β 
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware
}}

Latest revision as of 18:05, 10 November 2020

Redirect to: