| {{MitoPedia
| | #REDIRECT [[OroboPOS-Holder]] |
| |description=
| |
| The OroboPOS-Holder (POS-Holder), made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding the OroboPOS-Connector with the OroboPOS sensor head (POS) to the O2k-Chamber, and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-Chamber with the OroboPOS-Seal Tip. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.
| |
| Β | |
| Two units of this item are standard components mounted on the O2k-Main Unit.
| |
| |info=[[OroboPOS-Holder]]
| |
| }}
| |
| __TOC__
| |
| Β | |
| {{Template:Technical support integrated}}
| |
| == Assembly ==
| |
| Β | |
| ::::* Manual: [https://wiki.oroboros.at/images/2/29/MiPNet22.11_O2k-FluoRespirometer-manual.pdf MiPNet22.11_O2k-FluoRespirometer-manual.pdf] Section 5.2.
| |
| Β | |
| ::::* Video: [https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]
| |
| == A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber ==
| |
| Β | |
| :::* The OroboPOS-Holder may be damaged:
| |
| Β | |
| ::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (''see center picture below'').Β
| |
| ::::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]].
| |
| Β | |
| ::::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see'' pictures below).
| |
| Β | |
| [[File:OroboPOS-Holder002.jpg|left|200px]]
| |
| [[File:OroboPOS-Holder003.jpg|right|200px]]
| |
| [[File:OroboPOS-Holder004.jpg|centre|200px]]
| |
| Β | |
| * Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
| |
| * Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged.
| |
| * Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.
| |
| Β | |
| Β | |
| ::::* Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: [https://www.oroboros.at/index.php/o2k-products/spares/ Spare parts & consumables]
| |
| ''See'' [[Talk:O2k-OroboPOS-Holder MitoPedia|Discussion page]]
| |
| Β | |
| {{MitoPedia O2k and high-resolution respirometry
| |
| |mitopedia O2k and high-resolution respirometry=O2k hardware
| |
| }}
| |