== Assembly ==


::::* Manual: [https://wiki.oroboros.at/images/2/29/MiPNet22.11_O2k-FluoRespirometer-manual.pdf MiPNet22.11_O2k-FluoRespirometer-manual.pdf] Section 5.2.
::::* Video: [https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]
== A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber ==
== A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber ==


:::* The OroboPOS-Holder may be damaged:
:::* The OroboPOS-Holder may be damaged:


::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (''see center picture below''). - ''Follow the details in the video'' - Β»[https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]Β«
::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (''see center picture below'').Β  Β 
::::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]].
::::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]].


::::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see pictures below'').
::::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see'' pictures below).


[[File:OroboPOS-Holder002.jpg|left|200px]]
[[File:OroboPOS-Holder002.jpg|left|200px]]

Revision as of 13:48, 5 March 2020


high-resolution terminology - matching measurements at high-resolution


O2k-OroboPOS-Holder MitoPedia

Description

The OroboPOS-Holder (POS-Holder), made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding the OroboPOS-Connector with the OroboPOS sensor head (POS) to the O2k-Chamber, and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-Chamber with the OroboPOS-Seal Tip. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.

Two units of this item are standard components mounted on the O2k-Main Unit.


Reference: OroboPOS-Holder


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Assembly

A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber

  • The OroboPOS-Holder may be damaged:
  1. During chamber assembly, when fixing the chamber in a wrong rotational position with the O2k-Chamber Holder (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (see center picture below).
  2. After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the O2k-Main Unit.
  • We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (see pictures below).
OroboPOS-Holder002.jpg
OroboPOS-Holder003.jpg
OroboPOS-Holder004.jpg
  • Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
  • Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged.
  • Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.


See Discussion page