Difference between revisions of "O2k-OroboPOS-Holder MitoPedia"
From Bioblast
Line 9: | Line 9: | ||
{{Template:Technical support integrated}} | {{Template:Technical support integrated}} | ||
== Assembly == | |||
::::* Manual: [https://wiki.oroboros.at/images/2/29/MiPNet22.11_O2k-FluoRespirometer-manual.pdf MiPNet22.11_O2k-FluoRespirometer-manual.pdf] Section 5.2. | |||
::::* Video: [https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber] | |||
== A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber == | == A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber == | ||
:::* The OroboPOS-Holder may be damaged: | :::* The OroboPOS-Holder may be damaged: | ||
::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (''see center picture below''). | ::::# During chamber assembly, when fixing the chamber in a wrong rotational position with the [[O2k-Chamber Holder]] (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (''see center picture below'').Β Β | ||
::::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]]. | ::::# After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the [[O2k-Main Unit]]. | ||
::::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see pictures below | ::::* We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (''see'' pictures below). | ||
[[File:OroboPOS-Holder002.jpg|left|200px]] | [[File:OroboPOS-Holder002.jpg|left|200px]] |
Revision as of 13:48, 5 March 2020
- high-resolution terminology - matching measurements at high-resolution
O2k-OroboPOS-Holder MitoPedia
Description
The OroboPOS-Holder (POS-Holder), made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding the OroboPOS-Connector with the OroboPOS sensor head (POS) to the O2k-Chamber, and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-Chamber with the OroboPOS-Seal Tip. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.
Two units of this item are standard components mounted on the O2k-Main Unit.
Reference: OroboPOS-Holder
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Assembly
- Manual: MiPNet22.11_O2k-FluoRespirometer-manual.pdf Section 5.2.
A damaged OroboPOS-Holder may cause medium to leak from the O2k-Chamber
- The OroboPOS-Holder may be damaged:
- During chamber assembly, when fixing the chamber in a wrong rotational position with the O2k-Chamber Holder (from above) before the OroboPOS-Holder is fully inserted, and subsequently screwing tightly the OroboPOS-Holder (see center picture below).
- After fully unscrewing the OroboPOS-Holder and forcing during reassembly an improperly aligned holder into the thread of the O2k-Main Unit.
- We recommend to check before any reassembly the mechanical integrity of the OroboPOS-Holder (see pictures below).
- Left picture: brand new OroboPOS-Holder, no damage on the thread, no damage on the edge.
- Center picture: used OroboPOS-Holder, almost no damage on the thread, the edge is slightly damaged.
- Right picture: heavly used OroboPOS-Holder, the thread is slightly damaged, serious damage on the edge of the aperture.
- Replacement of a damaged OroboPOS-Holder is recommended. For ordering spare parts: Spare parts & consumables
See Discussion page
MitoPedia O2k and high-resolution respirometry:
O2k hardware