== O2k-OroboPOS-Holder: Leaky chamber due to damaged OroboPOS-Holder==
== O2k-OroboPOS-Holder: Leaky chamber due to damaged OroboPOS-Holder==


:::* In order not to experience a leakage after chamber reassembly, make sure that all the parts are in good conditions, especially the OroboPOS-Holder Β 
:::* In order not to experience a leakage after chamber reassembly, make sure that all the parts are in good conditions, especially the OroboPOS-Holder .
:::* The OroboPOS-Holder may experience some damage due to the not proper alignment with the thread Β 
:::* The OroboPOS-Holder may experience some damage due to the not proper alignment with the thread.
:::* The OroboPOS-Holder may experience some damage due to the not proper alignment with the chamber
:::* The OroboPOS-Holder may experience some damage due to the not proper alignment with the chamber.
:::* The OroboPOS-Holder may experience some damage due to frequent and bad performed mounting and dismounting procedures
:::* The OroboPOS-Holder may experience some damage due to frequent and bad performed mounting and dismounting procedures.


:::* In case your OroboPOS-Holder already presents dents on the surfaces ( see pictures below) you may experience leakage, sanding the edge of the OroboPOS-Holder with sand paper may solve the issue.
:::* In case your OroboPOS-Holder already presents dents on the surfaces ( see pictures below) you may experience leakage. Although sanding the edge of the OroboPOS-Holder with sand paper may solve the problem, replacement of a damaged OroboPOS-Holder is recommended.


[[File:OroboPOS-Holder002.jpg|left|200px]]
[[File:OroboPOS-Holder002.jpg|left|200px]]
Line 28: Line 23:
[[File:OroboPOS-Holder004.jpg|centre|200px]]
[[File:OroboPOS-Holder004.jpg|centre|200px]]


* Left picture: brand new OroboPOS-Holder, no damage on the thred, no damage on the edge
* Left picture: brand new OroboPOS-Holder, no damage on the thred, no damage on the edge.
* Center picture: used OroboPOS-Holder, almost no damage on the thred, the edge is slightly damaged Β 
* Center picture: used OroboPOS-Holder, almost no damage on the thred, the edge is slightly damaged.
* Right picture: heavly used OroboPOS-Holder, the thred is slightly damaged, serious damage on the edge of the aperture
* Right picture: heavly used OroboPOS-Holder, the thred is slightly damaged, serious damage on the edge of the aperture.


:::* For more information, please see: [https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]
:::* ''More details are shown in this video'' - Β»[https://www.youtube.com/watch?v=1MSsI1ePMu8&feature=youtu.be Chamber service Insert O2k Chamber]Β«


{{MitoPedia concepts}}
{{MitoPedia methods}}
{{MitoPedia O2k and high-resolution respirometry}}
{{MitoPedia topics}}
{{MitoPedia O2k and high-resolution respirometry
{{MitoPedia O2k and high-resolution respirometry
|mitopedia O2k and high-resolution respirometry=O2k hardware
|mitopedia O2k and high-resolution respirometry=O2k hardware
}}
}}

Revision as of 19:29, 4 March 2020


high-resolution terminology - matching measurements at high-resolution


O2k-OroboPOS-Holder MitoPedia

Description

The OroboPOS-Holder (POS-Holder), made from blue POM, is screwed into the copper block of the O2k-Main Unit, guiding the OroboPOS-Connector with the OroboPOS sensor head (POS) to the O2k-Chamber, and keeping the OroboPOS-Connector in a fixed position for sealing the O2k-Chamber with the OroboPOS-Seal Tip. In addition, the OroboPOS-Holder fixes the O2k-Chamber in an accurate rotational position by pressing against the angular cut of the glass chamber.

Two units of this item are standard components mounted on the O2k-Main Unit.


Reference: OroboPOS-Holder


Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



O2k-OroboPOS-Holder: Leaky chamber due to damaged OroboPOS-Holder

  • In order not to experience a leakage after chamber reassembly, make sure that all the parts are in good conditions, especially the OroboPOS-Holder .
  • The OroboPOS-Holder may experience some damage due to the not proper alignment with the thread.
  • The OroboPOS-Holder may experience some damage due to the not proper alignment with the chamber.
  • The OroboPOS-Holder may experience some damage due to frequent and bad performed mounting and dismounting procedures.
  • In case your OroboPOS-Holder already presents dents on the surfaces ( see pictures below) you may experience leakage. Although sanding the edge of the OroboPOS-Holder with sand paper may solve the problem, replacement of a damaged OroboPOS-Holder is recommended.
OroboPOS-Holder002.jpg
OroboPOS-Holder003.jpg
OroboPOS-Holder004.jpg
  • Left picture: brand new OroboPOS-Holder, no damage on the thred, no damage on the edge.
  • Center picture: used OroboPOS-Holder, almost no damage on the thred, the edge is slightly damaged.
  • Right picture: heavly used OroboPOS-Holder, the thred is slightly damaged, serious damage on the edge of the aperture.


MitoPedia O2k and high-resolution respirometry: O2k hardware