'''Problem'''
: In files measured with DatLab version 5.1.1.7 the default name for the "Amp" channel (amperometric channel, used also for fluorescence measurements) is "pX" instead of "Amp".
When data are exported with [Marks] / [Statistics], the default units (µM, and pmol/(ml s)) are displayed correctly, but the channel should be "Amp". This error depends on the version '''DatLab 5.1.1.7''' used for recording the file, not on the DatLab version used for viewing and further analysis.
'''Solution''':# Upgrade to DatLab 6 or higher, available [http://www.oroboros.at/?login on the OROBOROS homepage](access with your user code and password).
:# Files recorded with DatLab 5.1.1.7:  Rename the Amp channel, using the calibration window as described in [[DatLab#New_Features_in_DatLab_5.1|DatLab - new features]]. You may rename the "Amp" channel to any name you choose (e.g. Fluorescence, H2O2, ..).


[[Category:OroboPedia]]
[[Category:OroboPedia]]

Revision as of 10:38, 16 June 2015

Publications in the MiPMap
O2k-Manual
O2k calibration by DatLab 5. »Bioblast pdf«

» Versions

OROBOROS (2014-03) Mitochondr Physiol Network

Abstract: Gnaiger E (2014) O2k calibration by DatLab 5. Mitochondr Physiol Network 19.01(D01):41-8.

Updated version: »DatLab 6«


O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: