Difference between revisions of "MiPNet19.01D O2k-Calibration"
From Bioblast
Line 7: | Line 7: | ||
|abstract=Gnaiger E (2014) O2k calibration by DatLab 5. Mitochondr Physiol Network 19.01(D01):41-8. | |abstract=Gnaiger E (2014) O2k calibration by DatLab 5. Mitochondr Physiol Network 19.01(D01):41-8. | ||
: ''Updated version:'' '''»[[MiPNet19.18D O2k-Calibration |DatLab 6]]«''' | : ''Updated version:'' '''»[[MiPNet19.18D O2k-Calibration |DatLab 6]]«''' | ||
|mipnetlab=AT_Innsbruck_OROBOROS | |mipnetlab=AT_Innsbruck_OROBOROS | ||
}} | |||
{{Labeling | |||
|additional=Archive | |||
}} | }} | ||
== Amp channel displayed as pX == | == Amp channel displayed as pX == |
Revision as of 15:50, 7 June 2015
O2k calibration by DatLab 5. »Bioblast pdf« |
» Versions
OROBOROS (2014-03) Mitochondr Physiol Network
Abstract: Gnaiger E (2014) O2k calibration by DatLab 5. Mitochondr Physiol Network 19.01(D01):41-8.
- Updated version: »DatLab 6«
• O2k-Network Lab: AT_Innsbruck_OROBOROS
Labels:
Archive
Amp channel displayed as pX
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Problem
- In files measured with DatLab version 5.1.1.7 the default name for the "Amp" channel (amperometric channel, used also for fluorescence measurements) is "pX" instead of "Amp".
When data are exported with [Marks] / [Statistics], the default units (µM, and pmol/(ml s)) are displayed correctly, but the channel should be "Amp". This error depends on the version DatLab 5.1.1.7 used for recording the file, not on the DatLab version used for viewing and further analysis.
Solution:# Upgrade to DatLab 6 or higher, available on the OROBOROS homepage(access with your user code and password).
- Files recorded with DatLab 5.1.1.7: Rename the Amp channel, using the calibration window as described in DatLab - new features. You may rename the "Amp" channel to any name you choose (e.g. Fluorescence, H2O2, ..).