The light intensity of the LED (LED-intensity) and the signal amplification (Gain) can be adjusted in a wide range. The table suggests initial values, which can be optimised for specific applications.
  • The settings depend on the concentration of the fluorophore, which vary between different applications. Therefore, only recommendations for specific fluorophore concentrations are given. In the Amplexยฎ UltraRed assay the fluorophore is formed during the experiment.
  • The recommendations apply to experiments at 37 ยฐC. The Fluorescence intensity increases strongly at lower temperatures. Then the light intensity is reduced to avoid off-scale signals.
Application Sensor Filter set O2k output Light intensity (Amp polarization voltage) Gain Comment
AmplexTM UltraRed Fluorescence-Sensor Green or Smart Fluo-Sensor Green AmR Type II 100 - 500 1000 at c(AmR) = 10 ยตM
TMRM Fluorescence-Sensor Green or Smart Fluo-Sensor Green AmR Type III 100 - 500 1000 at c(TMRM) = 2 ยตM
Safranin Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue Saf Type III 100 - 250 1000 at c(safranin) = 2 ยตM
Rhodamine 123 Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue Saf Type III 100 - 250 1000 at c(Rh123) = 1 ยตM
Magnesium Greenโ„ข Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue MgG / CaG Type I and II 100 - 500 1000 at c(Mg Green) = 2 ยตM
Calcium Greenโ„ข Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue MgG / CaG Type I 300 - 500 1000 at c(Ca Green) = 2 ยตM

In the O2k-LED2-Module (Series D-G) polarization voltage is used, while in the O2k-Fluo Smart Module (Series H-I) fluorescence intensity. To use the same setting in the O2k-Fluo LED2-Module as in the O2k-Fluo Smart-Module use the following Excel sheet:

Download2.png Download the EXCEL calculation-sheet to adjust fluorescence intensity