Application Sensor Filter set O2k output Light intensity (Amp polarization voltage) Gain Comment
AmplexTM UltraRed Fluorescence-Sensor Green or Smart Fluo-Sensor Green AmR Type II 100 - 500 1000 at c(AmR) = 10 µM
TMRM Fluorescence-Sensor Green or Smart Fluo-Sensor Green AmR Type III 100 - 500 1000 at c(TMRM) = 2 µM
Safranin Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue Saf Type III 100 - 250 1000 at c(safranin) = 2 µM
Rhodamine 123 Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue Saf Type III 100 - 250 1000 at c(Rh123) = 1 µM
Magnesium Green™ Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue MgG / CaG Type I and II 100 - 500 1000 at c(Mg Green) = 2 µM
Calcium Green™ Fluorescence-Sensor Blue or Smart Fluo-Sensor Blue MgG / CaG Type I 300 - 500 1000 at c(Ca Green) = 2 µM

In the O2k-LED2-Module (Series D-G) polarization voltage is used, while in the O2k-Fluo Smart Module (Series H-I) fluorescence intensity. To use the same setting in the O2k-Fluo LED2-Module as in the O2k-Fluo Smart-Module use the following Excel sheet:

Download2.png Download the EXCEL calculation-sheet to adjust fluorescence intensity