Talk:MiPNet17.05 O2k-Fluo LED2-Module

From Bioblast

Titration and settings

  • 10 Β΅l AmR - 5 Β΅M final conc. in chamber
  • 4Β΅l HRP (500U/ml) - 1 U/ml final conc. in chamber* light intensity of LEDs: 1 mA ( DatLab: Amp polaristion voltage 100 mV) / Gain 1000
  • H2O2 calibration with a 15.8 Β΅M H2O2 solution - 2012-06-20
commercial stock: 880 mM
dilution 1 (1:166.67): 150 Β΅l in 25 ml H2O = 5.28 mM
dilution 2 (1:333.33): 150 Β΅l in 50 ml H2O + 500 Β΅l 1 mM HCl = 15.8 Β΅M
β†’ calibration: 2 * 5 Β΅l in 2 ml chamber = 2* 0.0395 Β΅M = total 0.079 Β΅M (= 79 nM)
  • new H2O2 calibration with a 28.16 Β΅M H2O2 solution, valid from from 2012-06-22:
commercial stock: 880 mM
dilution 1 (1:125): 200 Β΅l in 25 ml H2O = 7.04 mM
dilution 2 (1:250): 200 Β΅l in 50 ml H2O + 500 Β΅l 1 mM HCl = 28.16 Β΅M
β†’ calibration: 2 * 5 Β΅l in 2 ml chamber = 2* 0.0704 Β΅M = total 0.1408 Β΅M (= 140.8 nM)
Cookies help us deliver our services. By using our services, you agree to our use of cookies.