Line 1: Line 1:
=Titration and settings=
* 10 ยตlย  AmR - 5 ยตM final conc. in chamber
* 4ยตl HRP (500U/ml) - 1 U/ml final conc. in chamber* light intensity of LEDs: 1 mA ( DatLab: Amp polaristion voltage 100 mV) / Gain 1000
* H<sub>2</sub>O<sub>2</sub> calibration with a''' 15.8 ยตM H<sub>2</sub>O<sub>2</sub>''' solution - '''2012-06-20'''
:: '''commercial stock:''' 880 mM
:: dilution 1 (1:166.67): 150 ยตl in 25 ml H<sub>2</sub>O = 5.28 mM
:: dilution 2 (1:333.33): 150 ยตl in 50 ml H<sub>2</sub>O + 500 ยตl 1 mM HCl = '''15.8 ยตM'''
:: โ†’ '''calibration:''' 2 * 5 ยตl in 2 ml chamber = 2* 0.0395 ยตM = total 0.079 ยตM (= 79 nM)


* new H<sub>2</sub>O<sub>2</sub> calibration with a '''28.16 ยตM''' H<sub>2</sub>O<sub>2</sub> solution, valid from from '''2012-06-22''':
:: '''commercial stock:''' 880 mM
:: dilution 1 (1:125): 200 ยตl in 25 ml H<sub>2</sub>O = 7.04 mM
:: dilution 2 (1:250): 200 ยตl in 50 ml H<sub>2</sub>O + 500 ยตl 1 mM HCl = '''28.16 ยตM'''
:: โ†’ '''calibration''': 2 * 5 ยตl in 2 ml chamber = 2* 0.0704 ยตM = total 0.1408 ยตM (= 140.8 nM)