The OROBOROS Oxygraph-2k with DatLab software is world-wide the only instrument (sole source) which allows routine measurements to be made with specifications summarized under the term high-resolution respirometry (HRR).
#REDIRECT [[MiPNet06.05 Test Experiments on O2k-Specifications]]
<br />
<br />
Β 
== HRR ==
<br />
The limit of detection of oxygen flux is as low as 1 pmol O2.s-1.ml-1.Β  Β  Oxygen backdiffusion at zero oxygen is <4 pmol.s-1.ml-1. Oxygen consumption by the polarographic oxygen sensor at air saturation and standard barometric pressure (100 kPa) is 2.7 Β± 0.9 SD (114 test runs at 37 Β°C).Β  These highly standardized instrumental background fluxes are a linear function of oxygen concentration, which is used for automatic background correction of oxygen flux (DatLab).Β  Typical exponential time constants of the oxygen sensors are <4 s, used for dynamic corrections in kinetic studies. Signal noise at zero oxygen concentration is <0.05 Β΅M O2.
<br />
Continuous control of the OROBOROS Oxygraph-2k specifications are a fundamental component of [[Quality Assurance| Quality Assurance]] and distinguish high-resolution respirometry from any conventional oxygraph system and from non-quantitative approaches.
<br />
<br />
Β 
== Oxygraph-2k: technical specifications ==
=== General ===
Β 
* Dimensions - O2k-Main Unit: L 45 cm, W 31 cm, H 25 cm; 14 kg without packing box or Peli Case.
Β 
* Chambers: chemically inert and minimum oxygen diffusion:
Β 
::: - Two Duran glass chambers with 2 ml (1.5 to 3.5 ml) volume; 16 mm inner diameter.
Β 
::: - PVDF or PEEK stoppers with Viton O-rings.
Β 
::: - Electromagnetic stirrers with variable speed (100 to 900 rpm).
Β 
::: - PVDF- or PEEK-coated stirrer bars (6 mm diameter).
Β 
::: - Polarographic oxygen sensors (OroboPOS) sealed with butyl rubber seal tips.
Β 
* Built-in electronically regulated Peltier thermostat:
Β 
::: - Temperature range:
:::::{| class="wikitable"
|-
| At room temperature
| 4 Β°C to 47 Β°CΒ 
|-
| At lower ambient temperature
| 2 Β°C
|}
Β 
::: - Temperature stability:
:::::{| class="wikitable"
|-
| Temperature stability
| Β±0.002 Β°C over 90 min.
|}
Β 
::: - Temperature range:
:::::{| class="wikitable"
|-
| in 15 min
| 20 Β°C to 30 Β°CΒ 
|-
| in 20 min
| 30 Β°C to 20 Β°C
|}
Β 
* Data output through USB:
Β 
::: - Two oxygen signals.
Β 
::: - Pressure transducer, for absolute barometric pressure signal, resolution 0.1 kPa.
Β 
::: - Thermostat temperature signal, resolution 0.001 Β°C.
Β 
::: - O2k-MultiSensor: two additional amperometric signals (O2k-fluorescence; NO sensors), and two additional potentiometric signals pX: for TPP or pH).
Β 
*DatLab software:
Β 
::: - Display: simultaneous display of oxygen concentration and oxygen flux (negative slope or time derivative), together with any selected O2k-MultiSensor channels (raw or calibrated signals and corresponding time derivatives). Barometric pressure for oxygen calibration; temperature stability control.
Β 
::: - Automatic calibrations: oxygen and O2k-MultiSensors.
Β 
Β 
=== Specifications for high-resolution respirometry ===
Β 
* Oxygen signal:
Β 
::: - Noise at zero oxygen: <0.005 kPa (SD, 100 data points recorded at 1 s intervals) without smoothing (Β±0.003 kPa typical)
Β 
::: - Noise at air saturation: <0.010 kPa (SD, 100 data points recorded at 1 s intervals) without smoothing, at partial oxygen pressure of 20 kPa (Β±0.005 kPa typical)
Β 
::: - Time constant: <7 s at 25 Β°C (<4 s typical)
Β 
::: - O2 range of linearity: Oxygen partial pressure of 0 to 100 kPa
Β 
* Oxygen flux:
Β 
::: - Limit of detection: 0.5 pmol.s-1.ml-1 at steady-state over 5 min.
Β 
::: - Sensitivity (normoxic): <1 pmol.s-1.ml-1 at steady-state over 5 min at 20 - 40 Β°C.
Β 
::: - Sensitivity (hyperoxic): <3 pmol.s-1.ml-1 at steady-state over 5 min at 20 - 40 Β°C.
Β 
::: - Noise: <0.2 pmol.s-1.ml-1 after standard smoothing (120 s).
Β 
::: - O2 range of measurement: Flux measured at oxygen partial pressure up to 100 kPa and to <0.01 kPa based on DatLab analysis of oxygen kinetics (mitochondria and cells)
Β 
Β 
* Instrumental background for linear correction over the entire oxygen range:
Β 
::: - O2 backdiffusion at 0 kPa:
Β 
::::::::: <3 pmol.s-1.ml-1 at 20 to 40 Β°C (2.5 pmol.s-1.ml-1 typical)
Β 
::: - O2 consumption at 20 kPa:
Β 
::::::::: <4 pmol.s-1.ml-1 at 37 Β°C (3 pmol.s-1.ml-1 typical)
Β 
::::::::: <3 pmol.s-1.ml-1 at 25 Β°C (2 pmol.s-1.ml-1 typical).