Raw signal of the oxygen sensor: Difference between revisions

From Bioblast
mNo edit summary
mNo edit summary
Line 20: Line 20:
The raw signal is converted to oxygen concentration [ยตM or nmol/ml] or partial oxygen pressure [kPa] by calibration of the POS [[http://www.oroboros.at/index.php?o2k-o2calibration MiPNet12.08]].ย  Each gain requires a separate calibration of the oxygen signal. Therefore, after changing the gain you should not only perform a calibration at air saturation (as always before an experiment), but also a new zero oxygen calibration.ย  ย 
The raw signal is converted to oxygen concentration [ยตM or nmol/ml] or partial oxygen pressure [kPa] by calibration of the POS [[http://www.oroboros.at/index.php?o2k-o2calibration MiPNet12.08]].ย  Each gain requires a separate calibration of the oxygen signal. Therefore, after changing the gain you should not only perform a calibration at air saturation (as always before an experiment), but also a new zero oxygen calibration.ย  ย 


{{#set:Technical service=sensor|Technical service=gain}}
{{#set:Technical service=Sensor|Technical service=Gain}}
__SHOWFACTBOX__
__SHOWFACTBOX__
{{Troubleshooting}}
{{Troubleshooting}}

Revision as of 07:32, 24 September 2010

For trouble shooting purposes of the polarographic oxygen sensor (POS), for choosing an appropriate gain setting (default is Gain 2), but also to understand how the O2k actually works, it is helpful to observe the raw signal (on-line or off-line).

Observing the Raw Signal

Depending on your DatLab version, a predefined layout called โ€žB Trouble shootingโ€œ is in the โ€žGraph layoutโ€œ menu. If not, you can easily generate this Graph layout. To do so, take the following steps in DatLab [MiPNet12.07]:

  • From the main menu select โ€œLayoutโ€ menu and choose โ€ž1 Calibration Exp. G3-Tempโ€œ
  • From the main menu select โ€žGraphโ€œ / โ€žSelect Plotsโ€œ, by default you will see the โ€žplotsโ€œ tab from the drop down menus for the oxygen signals. Select โ€žO2 raw signalโ€œ instead of โ€žO2 concentrationโ€œ for both chambers.
  • Select the โ€žscalingโ€œ tabs: adjust the scaling for the oxygen signal to 0 โ€“ 10 Volt (Start at 0, Range 10)
  • At the bottom of the window insert a name for the new layout (โ€žB Trouble Shootingโ€œ); press โ€žSave Graph Layoutโ€œ.

Remember to select the period of the experiment your are interested in or select โ€œGraphโ€ / โ€œAutoscale Time Axisโ€ to view the entire experiment. The maximum signal is 10 volt. Above 9.99 V, the amplifier is in the saturation range, the apparent signal will remain constant at this value, and the flux will appear to be zero. The raw signal can be modified by choosing an appropriate gain setting. We recommend a gain of 2 as a default for experiments at and below air saturation, to avoid reaching the upper 10 volt limit without noticing. Typical recorded voltages at air saturation, in pure water or experimental media, should be

  • at Gain 8: offscale ( โ€œ9.99Vโ€)
  • at Gain 4 : ca. 8 - 10 V (do not use at air saturation, especially if your lab is situated at or near sea level)
  • at Gain 2: ca. 4 - 5 V
  • at Gain 1: ca. 2 V

The actual value of the raw signal at air saturation (during calibration) depends on experimental temperature, barometric pressure, and the sensitivity of the individual POS. At or near sea level higher values are recorded than at higher altitudes. The raw signal is converted to oxygen concentration [ยตM or nmol/ml] or partial oxygen pressure [kPa] by calibration of the POS [MiPNet12.08]. Each gain requires a separate calibration of the oxygen signal. Therefore, after changing the gain you should not only perform a calibration at air saturation (as always before an experiment), but also a new zero oxygen calibration.


Template:Troubleshooting

Cookies help us deliver our services. By using our services, you agree to our use of cookies.