Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Oxygen flux - instrumental background

From Bioblast
Revision as of 16:57, 19 August 2010 by Wiethuechter Anita (talk | contribs) (Created page with "Instrumental background oxygen flux http://www.oroboros.at/index.php?id=backgroundcorrection MiPNet14.06 in a respirometer is due to oxygen consumption by the POS, and ox...")
(diff) ← Older revision | Latest revision (diff) | Newer revision β†’ (diff)
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Instrumental background oxygen flux [MiPNet14.06] in a respirometer is due to oxygen consumption by the POS, and oxygen diffusion into or out of the aqueous medium in the O2k-chamber. It is measured in the range of experimental oxygen levels by a standardized instrumental background test, and is a property of the instrumental system. Instrumental background correction eliminates errors by systemic flux compensation, automatically performed by DatLab.

References

  • Gnaiger E (2008) Polarographic oxygen sensors, the oxygraph and high-resolution respirometry to assess mitochondrial function. In: Mitochondrial Dysfunction in Drug-Induced Toxicity (Dykens JA, Will Y, eds) John Wiley: 327-352. [pdf].




high-resolution terminology - matching measurements at high-resolution


Oxygen flux - instrumental background

Description