Difference between revisions of "OroboPOS-Connector"

From Bioblast
m
Line 21: Line 21:


::::# Unscrew the OroboPOS and check the electrical connections (the thread of the OroboPOS and the gold pin of the OroboPOS-Connector).
::::# Unscrew the OroboPOS and check the electrical connections (the thread of the OroboPOS and the gold pin of the OroboPOS-Connector).
::::# Remove any contaminating grease and moisture with a fine paper cloth.
::::# Remove any contaminating grease, salt crystals and moisture with a fine paper cloth.
::::# Moisten a paper cloth with 99.9 % ethanol and carefully clean the gold pin and threads. If the thread of the OroboPOS and the gold pin of the OroboPOS-Connector are discolored or appear stained, clean repeatedly with a paper towel moistened with 99.9 % ethanol until no more traces appear on the paper towel used for cleaning.
::::# If the thread of the OroboPOS and the gold pin of the OroboPOS-Connector are discolored or appear stained, moisten a paper cloth with 99.9 % ethanol and clean carefully until no more traces appear on the paper towel used for cleaning.
::::* If the O2k is used in a humid environment, the connector can be dried in an oven at up to 60 ยฐC for 1-2h. Follow the procedure with an [[Oxygen sensor test]] as a quality control of proper OroboPOS-Connector and OroboPOS performance.
::::* If the O2k is used in a humid environment, the connector can be dried in an oven at up to 60 ยฐC for 1-2h. Follow the procedure with an [[Oxygen sensor test]] as a quality control of proper OroboPOS-Connector and OroboPOS performance.



Revision as of 14:04, 3 December 2020


                  


O2k-Open Support

OroboPOS-Connector



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



Description

POS connector and cable connection.jpg

OroboPOS-Connector (blue POM), with male connection to OroboPOS sensor head (POS) and with cable and male plug fitting into O2k-Main Unit.


Reference: Product details and purchase information

Functions

The OroboPOS-Connector has several different functions:
  • It mechanically presses the POS against the glass chamber, ensuring a tight fit.
  • It connects the POS electronically to the Main unit transmitting the signal.
  • In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the Main unit. For this reason OroboPOS connectors for Series A-C and for Series D are NOT interchangeable.
The OroboPOS-Connector is sensitive to damage by ESD when disconnected from the O2k-Main Unit and when the POS head is removed. Further details: MiPNet14.01_ESD-damage.

Cleaning the body of the OroboPOS-Connector

When experimental media or electrolyte (during service of the polarographic oxygen sensor OroboPOS) leak into the mechanical part of the OroboPOS-Connector, operation of the spring mechanism may be restricted. In such cases, the body of the OroboPOS-Connector should be washed with distilled water. To do so, screw the OroboPOS head onto the OroboPOS-Connector. Make sure that the plug connecting to the O2k-Main Unit does not get wet (protect with parafilm). Rinse the body of the OroboPOS-Connector with distilled water, pulling the spring-loaded part at the base several times. Salt crystals and media clogging the mechanical part are thus washed out. Wipe off the water while again pulling down the spring-loaded part.

Cleaning the electrical connections

  1. Unscrew the OroboPOS and check the electrical connections (the thread of the OroboPOS and the gold pin of the OroboPOS-Connector).
  2. Remove any contaminating grease, salt crystals and moisture with a fine paper cloth.
  3. If the thread of the OroboPOS and the gold pin of the OroboPOS-Connector are discolored or appear stained, moisten a paper cloth with 99.9 % ethanol and clean carefully until no more traces appear on the paper towel used for cleaning.
  • If the O2k is used in a humid environment, the connector can be dried in an oven at up to 60 ยฐC for 1-2h. Follow the procedure with an Oxygen sensor test as a quality control of proper OroboPOS-Connector and OroboPOS performance.

Replacing the OroboPOS-Connector

If you need to replace the Orobo-POS Connector, an OroboPOS-Connector Service may be possible.