== Functions == | == Functions == | ||
The OroboPOS-Connector has several different functions: | ::: The OroboPOS-Connector has several different functions: | ||
* It mechanically presses the [[POS]] against the [[glass chamber]], ensuring a tight fit. | ::::* It mechanically presses the [[POS]] against the [[glass chamber]], ensuring a tight fit. | ||
* It connects the [[POS]] electronically to the [[ | ::::* It connects the [[POS]] electronically to the [[Main unit]] transmitting the signal. | ||
* In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the [[Main unit]]. For this reason OroboPOS connectors for Series A-C and for Series D are NOT interchangeable. | ::::* In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the [[Main unit]]. For this reason OroboPOS connectors for Series A-C and for Series D are NOT interchangeable. | ||
The OroboPOS-Connector is | :::: The OroboPOS-Connector is sensitive to damage by [[ESD]] when disconnected from the [[O2k-Main Unit]] and when the POS head is removed. Further details: [[MiPNet14.01_ESD-damage]]. | ||
=== Cleaning | === Cleaning === | ||
:::: When experimental media or electrolyte (during [[MiPNet19.18B POS-service |service of the polarographic oxygen sensor OroboPOS]] leak into the mechanical part of the OroboPOS-Connector, operation of the spring mechanism may be restricted. In such cases, the body of the OroboPOS-Connector should be washed with distilled water. | |||
To do so, screw the OroboPOS head onto the OroboPOS-Connector Make sure that the plug connecting to the O2k-Main Unit does not get wet (protect with parafilm). Rinse the body of the OroboPOS-Connector with distilled water, pulling the spring-loaded part at the base several times. Salt crystals and media clogging the mechanical part are thus washed out. Wipe off the water while again pulling down the spring-loaded part. | |||
=== Cleaning the electrical connections === | === Cleaning the electrical connections === | ||
Unscrew the | :::: Unscrew the OroboPOS and check the electrical connections (the gold pin and thread on the OroboPOS and OroboPOS-Connector). Remove any contaminating grease and moisture with a fine paper cloth. Soak a paper cloth in ethanol and carefully clean the gold pin and threads. If the gold pin and thread on the POS connector are discoloured or appear stained, clean repeatedly with a paper towel soaked in water and subsequently in ethanol, until no more traces appear on the paper towel used for cleaning. If the O2k is used in a humid environment, apply a drop of contact oil from the [[OroboPOS-Service_Kit|OroboPOS-Service Kit]] to each of the golden pins and threads on POS connector and POS. Perform an [[O2 sensor test]] as a quality control of proper OroboPOS-connector and sensor performance. | ||
Β | |||
Β Β | Β Β | ||
[[Category:Technical service]] | [[Category:Technical service]] |
Revision as of 08:49, 9 October 2016
OroboPOS-Connector
Description | OroboPOS-Connector (blue POM), with male connection to OroboPOS sensor head (POS) and with cable and male plug fitting into O2k-Main Unit.
Two units of this item are standard components of the O2k-Assembly Kit (O2k-Core), mounted on the OroboPOS-Holders at the O2k-Main Unit. |
---|---|
Product ID | 22200-01 |
Type | O2k, O2k-Assembly Kit, OroboPOS |
Link | O2k-Assembly Kit, O2k-Manual |
Image |
MitoPedia O2k and high-resolution respirometry:
O2k hardware
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Functions
- The OroboPOS-Connector has several different functions:
- It mechanically presses the POS against the glass chamber, ensuring a tight fit.
- It connects the POS electronically to the Main unit transmitting the signal.
- In O2k Series A-C it also performs the current to voltage conversion / amplification. For Series D and upwards this function has been incorporated into the Main unit. For this reason OroboPOS connectors for Series A-C and for Series D are NOT interchangeable.
- The OroboPOS-Connector has several different functions:
- The OroboPOS-Connector is sensitive to damage by ESD when disconnected from the O2k-Main Unit and when the POS head is removed. Further details: MiPNet14.01_ESD-damage.
Cleaning
- When experimental media or electrolyte (during service of the polarographic oxygen sensor OroboPOS leak into the mechanical part of the OroboPOS-Connector, operation of the spring mechanism may be restricted. In such cases, the body of the OroboPOS-Connector should be washed with distilled water.
To do so, screw the OroboPOS head onto the OroboPOS-Connector Make sure that the plug connecting to the O2k-Main Unit does not get wet (protect with parafilm). Rinse the body of the OroboPOS-Connector with distilled water, pulling the spring-loaded part at the base several times. Salt crystals and media clogging the mechanical part are thus washed out. Wipe off the water while again pulling down the spring-loaded part.
Cleaning the electrical connections
- Unscrew the OroboPOS and check the electrical connections (the gold pin and thread on the OroboPOS and OroboPOS-Connector). Remove any contaminating grease and moisture with a fine paper cloth. Soak a paper cloth in ethanol and carefully clean the gold pin and threads. If the gold pin and thread on the POS connector are discoloured or appear stained, clean repeatedly with a paper towel soaked in water and subsequently in ethanol, until no more traces appear on the paper towel used for cleaning. If the O2k is used in a humid environment, apply a drop of contact oil from the OroboPOS-Service Kit to each of the golden pins and threads on POS connector and POS. Perform an O2 sensor test as a quality control of proper OroboPOS-connector and sensor performance.