| The term "open chamber" is used to denote a situation in which the liquid phase is in equilibrium with the atmosphere, resulting in a net zero O2 [[flux at open chamber]].
| |
| To maintain signal stability and protection from contamination this is NOT done by leaving the chamber open (no stoppers) but by creating a defined gas phase in the chamber between liquid phase and stopper using the a [[stopper spacer]].
| |
| This procedure is described in detail in the chapter "Air calibration" in [[http://www.oroboros.at/index.php?id=o2k-o2calibration MiPNet12.08]].
| |
|
| |
| See also:
| |
|
| |
| [[Air calibration]]
| |
|
| |
| [[Closed chamber]]
| |
|
| |