Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "O2k-Titration Set"

From Bioblast
Line 5: Line 5:
|product type=O2k, O2k-Module, O2k-Core, Microsyringe, Catalogue
|product type=O2k, O2k-Module, O2k-Core, Microsyringe, Catalogue
|info=[[MiPNet09.12 O2k-Titrations]]
|info=[[MiPNet09.12 O2k-Titrations]]
|product image=[[File:Hamilton Syringes for Manual Titration.jpg|right|180px|link=]]
|product image=[[Image:Hamilton Syringes for Manual Titration.jpg|right|180px|link=http://www.bioblast.at/index.php/Titration_Set]]
}}
}}



Revision as of 11:38, 13 November 2013

Template:Mainpage orobopedia

                



O2k-Titration Set

O2k-Catalogue

Description The Titration Set consists of 6x 10 ยตl, 6x 25 ยตl, and 2x 50 ยตl Hamilton glass syringes, and 3 spare plungers for the 10 ยตl syringes. It is a basic HRR-Accessory included in the O2k-Core Unit.
Product ID 20820-01
Type O2k, O2k-Module, O2k-Core, Microsyringe, Catalogue
Link MiPNet09.12 O2k-Titrations
Image
Hamilton Syringes for Manual Titration.jpg


The Titration Set consists of

6 Microsyringe\10 mm3 51/0.13 mm 10 mm3
6 Microsyringe\25 mm3 51/0.15 mm 25 mm3
2 Microsyringe\50 mm3 51/0.15 mm 50 mm3
3 Plunger\10 mm3 for 10 mm3 syringe


Handling of the syringes

  • separate inhibitors from substrates
  • rinse the outside with water immediately after use
  • at the start of a day: rinse the outside with the pure solvent and wash the syringes 3 times with the pure solvent
  • rinse with pure solvent before use
  • between two runs during the day: rinse the outside with water
  • at the end of a day: wash 3 x with last used solvent, rinse 3 x with EtOH 100 %
  • storage: ethanol 100% (short), dry (longtime)