== Treatment of the syringes ==
== Handling of the syringes ==
* for long time storage of the syringes fill them with 100 % ethanol and store them in 15 ml Falcons filled with 100 % ethanol that the needle and the beginning of the glass syringe are covered with liquid
* separate inhibitors from substrates
* before using the syringes always rinse them outside with the solvent (water or ethanol) of the chemical
* rinse the outside with water immediately after use
* before using the syringes wash them 3 times with the solvent - be aware that the whole syringe is filled till the end of the plunger!
* at the start of a day: rinse the outside with the pure solvent and wash the syringes 3 times with the pure solvent
* between titration: rinse the outside with water (to get rid of the respiration medium) and if necessary with ethanol depending on the solvent of the chemical
* rinse with pure solvent before use
•Separate inhibitors from substrates
* between two runs during the day: rinse the outside with water
•Between two runs during the day: rinse outside with water
* at the end of a day: wash 3 x with last used solvent, rinse 3 x with EtOH 100 %
•End of the day: 3 x solvent, 3 x EtOH 100%
* storage: ethanol 100% (short), dry (longtime)
•storage: in 100 % ethanol

Revision as of 09:18, 12 February 2013

Template:Mainpage orobopedia

                



O2k-Titration Set

O2k-Catalogue

Description The Titration Set consists of 6x 10 µl, 6x 25 µl, and 2x 50 µl Hamilton glass syringes, and 3 spare plungers for the 10 µl syringes. It is a basic HRR-Accessory included in the O2k-Core Unit.
Product ID 20820-01
Type O2k, O2k-Module, O2k-Core, Microsyringe, Catalogue
Link MiPNet09.12 O2k-Titrations
Image
Hamilton Syringes for Manual Titration.jpg


The Titration Set consists of

6 Microsyringe\10 mm3 51/0.13 mm 10 mm3
6 Microsyringe\25 mm3 51/0.15 mm 25 mm3
2 Microsyringe\50 mm3 51/0.15 mm 50 mm3
3 Plunger\10 mm3 for 10 mm3 syringe


Handling of the syringes

  • separate inhibitors from substrates
  • rinse the outside with water immediately after use
  • at the start of a day: rinse the outside with the pure solvent and wash the syringes 3 times with the pure solvent
  • rinse with pure solvent before use
  • between two runs during the day: rinse the outside with water
  • at the end of a day: wash 3 x with last used solvent, rinse 3 x with EtOH 100 %
  • storage: ethanol 100% (short), dry (longtime)