|title=[[Image:O2k-Manual.jpg|right|70px|link=http://oroboros.at/?O2k-Manual|O2k-Manual]]Gnaiger E (2014) Service of the polarographic oxygen sensor OroboPOS. Mitochondr Physiol Network 19.01(B01): 19-24. | |title=[[Image:O2k-Manual.jpg|right|70px|link=http://oroboros.at/?O2k-Manual|O2k-Manual]]Gnaiger E (2014) Service of the polarographic oxygen sensor OroboPOS. Mitochondr Physiol Network 19.01(B01): 19-24. | ||
Line 19: | Line 20: | ||
|articletype=Protocol; Manual, MiPNet-online Publication | |articletype=Protocol; Manual, MiPNet-online Publication | ||
}} | }} | ||
Revision as of 09:44, 17 May 2014
Gnaiger E (2014) Service of the polarographic oxygen sensor OroboPOS. Mitochondr Physiol Network 19.01(B01): 19-24. |
Β» Open Access-MiPNet19.01B.pdf; Versions
OROBOROS (2014-03) Mitochondr Physiol Network
Abstract: O2k-Core Manual Chapter B
- >> O2k-Manual: Overall contents
- >> Product: OROBOROS Oxygraph-2k, O2k-Catalogue
β’ Keywords: HRR
β’ O2k-Network Lab: AT_Innsbruck_OROBOROS
Labels: MiParea: Instruments;methods
HRR: Oxygraph-2k, O2k-Manual
O2k-Core
- Frequency of POS-Service: Instead of a fixed interval we suggest to follow an Instrumental quality assurance plan and to perform a POS service only when the conditions listed in this manual (see above for link to pdf) are met.