Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "MiPNet19.18B POS-service"

From Bioblast
(28 intermediate revisions by 9 users not shown)
Line 1: Line 1:
{{OROBOROS header page name}}
{{Template:OROBOROS support page name}}
{{Publication
{{Publication
|title=[[Image:O2k-Manual.jpg|right|70px|link=O2k-Manual|O2k-Manual]] Service of the polarographic oxygen sensor OroboPOS. [[Media:MiPNet19.18B POS-Service.pdf |»Bioblast pdf«]]
|title=[[Image:O2k-Manual.jpg|right|70px|link=O2k-Manual|O2k-Manual]] Service of the polarographic oxygen sensor OroboPOS.  
|info=[http://www.bioblast.at/index.php/File:MiPNet19.18B_POS-Service.pdf Versions]
|info=[[File:PDF.jpg|100px|link=http://wiki.oroboros.at/images/6/6a/MiPNet19.18B_POS-Service.pdf |Bioblast pdf]] »[http://www.bioblast.at/index.php/File:MiPNet19.18B_POS-Service.pdf Versions] |authors=Oroboros
|authors=OROBOROS
|year=2021-06-23
|year=2015-11-02
|journal=Mitochondr Physiol Network
|journal=Mitochondr Physiol Network
|abstract=Gnaiger E (2015) Service of the polarographic oxygen sensor OroboPOS. Mitochondr Physiol Network 19.18(B03):17-22.
|abstract=Gnaiger E (2021) Service of the polarographic oxygen sensor OroboPOS. Mitochondr Physiol Network 19.18(B09):1-7.
{{MiPNet pdf page linking to MitoPedia}}
:» Product: [[OroboPOS]]
|mipnetlab=AT_Innsbruck_Oroboros
}}
 
{{Template:Keywords: Oxygen signal}}
 


:» [[O2k-Videosupport]]
:» '''O2k-SOP''': Polarographic oxygen sensors: calibration, accuracy and quality control SOP. [[MiPNet06.03 POS-Calibration-SOP |»POS-Calibration-SOP«]]
:» Product: [[OroboPOS]], [[Oxygraph-2k|O2k]], [[OROBOROS O2k-Catalogue |O2k-Catalogue]]
|keywords=HRR
|mipnetlab=AT_Innsbruck_OROBOROS
}}
{{Labeling
{{Labeling
|area=Instruments;methods
|area=Instruments;methods
|instruments=Oxygraph-2k, O2k-Manual
|instruments=Oxygraph-2k, O2k-Manual
|additional=O2k-Core
|additional=O2k-Core, DL6, DL7, DL6a7, SeriesG, SeriesH, SeriesI,
MitoFit2022QC
}}
}}
{{Template:Technical support integrated}}
== Sensor service by OROBOROS ==
The anode of the POS consists of a solid silver ring coated with  silver chloride (There is no "Ag coating"). The AgCl coating will GROW  during use. A typical contamination will be the formation of AgO. AgO  will be preferentially removed by the ammonia treatment during sensor  service, however part of the AgCl layer will also be removed. Any  defects in the AgCl layer are restored by running the O2k for a few  hours (or over night) after the sensor service or just by continued use.
We offer a sensor service in our workshop. This is essentially the same that you can do on your own. The main advantage of the sensor service in our workshop is the presence of a very experienced and very patient team, carrying out many iterations of NH3 cleaning. Running the O2k over night may be required to stabilize a newly serviced OroboPOS.

Revision as of 10:17, 5 May 2022


                  


O2k-Open Support

MiPNet19.18B POS-service


Publications in the MiPMap
O2k-Manual
Service of the polarographic oxygen sensor OroboPOS.

» Bioblast pdf »Versions

Oroboros (2021-06-23) Mitochondr Physiol Network

Abstract: Gnaiger E (2021) Service of the polarographic oxygen sensor OroboPOS. Mitochondr Physiol Network 19.18(B09):1-7.

O2k-technical support and open innovation
Open the pdf document above.
» Current O2k-series: NextGen-O2k Series XB and O2k Series J
» Current software versions DatLab 8.0: MitoPedia: DatLab
» Product: OroboPOS


O2k-Network Lab: AT_Innsbruck_Oroboros



Questions.jpg


Click to expand or collaps



Labels: MiParea: Instruments;methods 





HRR: Oxygraph-2k, O2k-Manual 

O2k-Core, DL6, DL7, DL6a7, SeriesG, SeriesH, SeriesI, MitoFit2022QC