|title=[[Image:O2k-Manual.jpg|right|70px|link=http://wiki.oroboros.at/index.php/O2k-Manual|O2k-Manual]] Electrostatic discharge (ESD): damage and protection.
|title=[[Image:O2k-Manual.jpg|right|70px|link=http://wiki.oroboros.at/index.php/O2k-Manual|O2k-Manual]] Electrostatic discharge (ESD): damage and protection.

Revision as of 13:42, 12 February 2015

Template:OROBOROS navigation line page name

Publications in the MiPMap
O2k-Manual
Electrostatic discharge (ESD): damage and protection.

Β» Open Access-MiPNet14.01.pdf; Versions

OROBOROS (2011) Mitochondr Physiol Network

Abstract: Fasching M (2011) Electrostatic discharge (ESD): damage and protection. Mitochondr Physiol Network 14.01(07).

>> Product: OROBOROS Oxygraph-2k, O2k-Catalogue

β€’ Keywords: HRR

β€’ O2k-Network Lab: AT Innsbruck OROBOROS


Labels: MiParea: Instruments;methods 





HRR: O2k-Manual 

O2k-Manual 

Further information

The best known experience of an electrostatic discharge (ESD) is feeling a spark when touching some object or another person. ESD is the sudden and momentary electric current that flows between two objects at different electrical potentials. Such a potential difference may commonly build up by

  • walking across the floor,
  • handling packaging material such as styrofoam and other plastic materials,
  • removing adhesive tape from a roll or container,
  • transporting equipment on trays or carts,
  • sliding items on a work bench.


General electrostatic discharge (ESD) protection procedures (most important: touching the oxygraph housing before touching any sensitive part) should be observed when handling any kind of electronic equipment, including the OROBOROS Oxygraph-2k.

Attention: This only works when the Oxygraph-2k is connected to the mains power supply. If the mains cable of the oxygraph is unplugged other grounded objects have to be found for touching (e.g. heating, water pipes).



Procedures that may be susceptible to damage components by ESD are especially:



  • Handling the POS connector with removed POS and / or when unplugged form the main unit.
  • Handling the pX channel: There are two situations when a ESD caused by the operators body could reach the electronics of the oxygraph while working with the pX channel:
    • Directly by touching the BNC port during plugging in or out an electrode.
    • Theoretically an electrical contact is also possible by touching the needles of the TIP syringes when any potentiometric sensor (TPP, pH, Ca,..) is in the chamber filled with medium (medium = electrolyte).



In these situations extra care should be taken to strictly follow the guidelines to prevent damage to the electronics (MiPNet14.01).