Cookies help us deliver our services. By using our services, you agree to our use of cookies. More information

Difference between revisions of "Dithionite"

From Bioblast
Line 3: Line 3:
|description=Zero oxygen solution powder, Na<sub>2</sub>S<sub>2</sub>O<sub>4</sub>, used for [[MiPNet19.18D O2k-calibration |calibration of oxygen sensors]] at zero oxygen, or for stepwise reduction of oxygen concentration in [[MiPNet14.06 InstrumentalO2Background |instrumental O2 background tests]].
|description=Zero oxygen solution powder, Na<sub>2</sub>S<sub>2</sub>O<sub>4</sub>, used for [[MiPNet19.18D O2k-calibration |calibration of oxygen sensors]] at zero oxygen, or for stepwise reduction of oxygen concentration in [[MiPNet14.06 InstrumentalO2Background |instrumental O2 background tests]].
|info=[[MiPNet06.03 POS-calibration-SOP]], [[MiPNet14.06 InstrumentalO2Background]]
|info=[[MiPNet06.03 POS-calibration-SOP]], [[MiPNet14.06 InstrumentalO2Background]]
}}
{{MitoPedia concepts|type=Chemicals
}}
}}
{{MitoPedia methods
{{MitoPedia methods
|mitopedia method=Respirometry
|mitopedia method=Respirometry
}}
{{MitoPedia O2k and high-resolution respirometry|type=Chemicals
}}
}}
== Preparation ==
== Preparation ==
Line 21: Line 17:




::::* From [[MiPNet06.03 POS-Calibration-SOP|MiPNet06.03 POS-Calibration-SOP (Version 14)]]: Prepare "zero solution" by dissolving c. 20 mg sodium hydrosulfite ([[OroboPOS-Service Kit]]) or two tips of a spatula in 0.5 ml of water. Mix in a small vial with minimum gas space. Use fresh. Dithionite may not work after prolonged storage.
::::* From [[MiPNet06.03 POS-calibration-SOP|MiPNet06.03 POS-calibration-SOP (Version 14)]]: Prepare "zero solution" by dissolving c. 20 mg sodium hydrosulfite ([[OroboPOS-Service Kit]]) or two tips of a spatula in 0.5 ml of water. Mix in a small vial with minimum gas space. Use fresh. Dithionite may not work after prolonged storage.


Β  '''Further information Β»[[Talk:Dithionite]]'''
Β  '''Further information Β»[[Talk:Dithionite]]'''

Revision as of 10:51, 14 August 2016


high-resolution terminology - matching measurements at high-resolution


Dithionite

Description

Zero oxygen solution powder, Na2S2O4, used for calibration of oxygen sensors at zero oxygen, or for stepwise reduction of oxygen concentration in instrumental O2 background tests.

Abbreviation: Na2S2O4

Reference: MiPNet06.03 POS-calibration-SOP, MiPNet14.06 InstrumentalO2Background


MitoPedia methods: Respirometry 

Preparation

Preparation of 10 mM Dithionite solution for Instrumental Background (dissolved in Phosphate buffer, 50 mM, pH 8)
  1. Weigh 0.017 g Dithionite, sodium hydrosulfite, and transfer to 10 mL volumetric glass flask.
  2. Adjust final volume to 10 ml with Phosphate buffer and keep closed to avoid decomposing.
Note: Sodium dithionite cannot be stored for a long period of time, because it is slowly oxidized by air when in solution. Even with the absence of air, solutions of sodium dithionite deteriorate. Therefore when Dithionite is open for a long time, we use a concentration of 30 mM for the Instrumental Background. The experiments showed the identical results with 10 and 30 mM Dithionite stock.