Volume-Calibration Ring sV (A or B), white PVDF, attached to PEEK O2k-Stopper sV.
Reference: Product details and purchase information
MitoPedia O2k and high-resolution respirometry:
O2k-Open Support
Technical support
- Volume calibration of the 0.5-mL O2k-chamber
- Loosen the stopper fixation rings (A) and (B) with the Allen key.
- Follow the instructions in the O2k-Manual: MiPNet22.11 O2k-FluoRespirometer manual
- Tighten the screws of the fixation rings (A) and (B) gently with the Allen key to ensure that the calibration ring cannot be displaced. Tightening the screw with too much force may damage the thread of the calibration ring. Check for a sufficient fixation by trying to rotate the calibration ring.
- Volume calibration of the 0.5-mL O2k-chamber
Keywords
- Β»Bioblast links: Chamber volume - >>>>>>> - Click on [Expand] or [Collapse] - >>>>>>>
- Β» Volume
- Units: Liter [L]; 1 L = 1 dm3 = 10-3 m3; 1 mL = 1 cm3; 1 Β΅L = 1 mm3
- Β» Volume
- On chamber volume
- Components of the experimental O2k-chambers
- Β» Stirrer-Bar\white PVDF\15x6 mm
- Β» Stirrer-Bar sV\white PVDF\11.5x6.2 mm
- Β» Stopper\black PEEK\conical Shaft\central Port
- Β» Stopper sV\black PEEK\conical Shaft\central Port
- Β» O-ring\Viton\12.5x1 mm
- Β» O-ring sV\Viton\9.5x1 mm
- Β» Volume-Calibration Ring
- Β» Volume-Calibration Ring sV
- Β» Stopper-Spacer
- Β» Cover-Slip\black
- Β» O2k-Chamber Holder
- Β» O2k-Chamber Holder sV
- Β» OroboPOS-Holder
- Β» OroboPOS-Holder sV
- Maintencance
MitoPedia O2k and high-resolution respirometry: O2k hardware, O2k-Open Support, O2k-Respirometry, O2k-FluoRespirometry