I have a question regarding NO-sensor calibration.. What method would you recommend?
I just tried with a solution (outside of the oroborous) recommended from the factory (0.05M H2SO4, 0.02M KI with addition of nitrite in different concentrations) This method worked fine, however, is this possible to do in the oroborous? Is that extremely acidic solution harmful to the POS membranes? Best regards Tomas Shiffer
Answer: Thats the method I always used. Its fine for the oxygraph (and of course very fine for the PTFE O2 membrane) but I have to admit the table I used got some "scars" :-) (from the KI).
Fasching Mario 11:45, 10 March 2011 (CET)