MiPNet06.03 POS-calibration-SOP

From Bioblast
Revision as of 11:59, 2 March 2016 by Gnaiger Erich (talk | contribs)

                



MiPNet06.03 POS-calibration-SOP

Publications in the MiPMap
O2k-Protocols
O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration.

» Bioblast pdf

OROBOROS (2016-02-09) Mitochondr Physiol Network

Abstract: Gnaiger E (2016) O2k Quality Control 1: Polarographic oxygen sensors and accuracy of calibration. Mitochondr Physiol Network 06.03(15):1-8. » Versions

O2k-Protocols SOP: High-resolution respirometry (HRR; OROBOROS Oxygraph-2k, O2k) critically depends on maintenance (MiPNet19.18B POS-Service) and accurate calibration of the polarographic oxygen sensors (OroboPOS, POS). Standard operating procedures (O2k-SOP) are described as part of the MitoFit Quality Control System. (1) Clean and ready-to-use O2k-Chambers are a basic requirement (MiPNet19.03). (2) Quality control for evaluation of proper POS function (SOP: O2 sensor test). (3) Accurate POS calibration (MiPNet19.18D O2k-Calibration). This is part 1 of O2k Quality Control.

» O2k-Protocols SOP: O2k Quality Control 2 »MiPNet14.06 InstrumentalO2Background«
» O2k-Manual »MiPNet19.18D_O2k-Calibration«
» Product: OroboPOS, O2k, O2k-Catalogue
» Template: O2k Calibration List


O2k-Network Lab: AT_Innsbruck_OROBOROS


Labels: MiParea: Instruments;methods 





HRR: Oxygraph-2k, O2k-Protocol 

O2k-SOP 



Template NextGen-O2k.jpg


MitoPedia O2k and high-resolution respirometry: O2k-Open Support 



O2 sensor test

Further details

O2k-QCS Next step - O2k Quality Control 2: »MiPNet14.06 InstrumentalO2Background«