OroboPOS-Connector

From Bioblast

Jump to: navigation, search
Description OroboPOS-Connector, with male connection to OroboPOS sensor head (POS) and with cable and male plug fitting into O2k-Main Unit.

O2k-Core: 2 in O2k-Assembly Kit, mounted on the OroboPOS-Holders at the O2k-Main Unit.

Product ID 22200-01
Type O2k, O2k-Assembly Kit, OroboPOS
Link Previous Product ID #25000-34, O2k-Assembly Kit, O2k-Manual @OROBOROS
Image
POS connector and cable connection.jpg


The two electronic connections of the POS connector (to the POS and to the Main unit) make a POS connector that has been disconnected from the POS or from the Main unit or from both particular sensitive to damage by ESD. Therefore it is of special importance to follow the guidelines for ESD protection [MiPNet14.1] whenever handling a POS connector.

Contents

Functions

The POS connector has several different functions:

Possible Problems

Nearly all problems usually associated with the POS may in fact be caused by the POS connectors, e.g.


Performing the Sensor Test

Confirm the problem by doing a Sensor test in the absence of biological material

Option 1.A: No problem was visible in the Sensor Test

check the following DatLab settings

If the problem is not visible when observing the Raw signal but is visible when observing the calibrated oxygen signal, then there is probable a problem with the POS calibration.

Option 1.B: The problem was visible in the Sensor Test

Localizing the Problem

Locate the Problem

Option 2.A: The problem was localized on the POS connector

Solutions 2.A Follow the instructions for cleaning and applying contact oil as described in the sensor service manual [MiPNet08.04] and in more detail at Cleaning the POS connector. If you have also done a Sensor service or changed the POS membranes allow the oxygraph to run over night before repeating the Sensor test. If all service precautions fail to solve the problem, or a new POS connector has to be applied.

Option 2.B: The problem was not localized on the POS connector

Solutions 2.B If the problem was localized on an other component follow the instructions for this component. If you could not locate the problem contact OROBOROS Instruments.

Facts about OroboPOS-ConnectorRDF feed
DescriptionOroboPOS-Connector, with male connection to OroboPOS sensor head (POS) and with cable and male plug fitting into O2k-Main Unit. O2k-Core: 2 in O2k-Assembly Kit, mounted on the OroboPOS-Holders at the O2k-Main Unit.
Product id22200-01  +
Product image
POS connector and cable connection.jpg
  +
Product typeO2k  +, O2k-Assembly Kit  +, and OroboPOS  +